Expanding Access to Retinal Imaging Through Patient-Operated Optical Coherence Tomography in a Veterans Affairs Retina Clinic

This study evaluated the feasibility, image quality, and referral accuracy of a patient-operated optical coherence tomography (OCT) device (SightSync) compared with technician-acquired Heidelberg OCT. This study was conducted in a Veterans Affairs retina clinic (Cleveland, Ohio), resulting in a predominantly male (98%) study population representative of the local veteran demographics. One hundred patients attempted self-administered OCT imaging after brief instruction, yielding 118 successful scans (59% of eyes) with no significant association between scan success and age, visual acuity, or diagnosis. Quantitative analysis of 142 paired images showed that SightSync produced interpretable scans with comparable sharpness to Heidelberg OCT, though signal- and intensity-based metrics (signal-to-noise ratio; SNR, contrast-to-noise ratio; CNR, entropy, pixel intensity; p90) were lower, consistent with hardware differences between a compact patient-operated prototype and a clinical-grade system. Among 121 high-quality SightSync scans, referral decisions demonstrated strong agreement with Heidelberg OCT, with a sensitivity of 83.9%, specificity of 75.6%, and a negative predictive value of 93.2%, indicating reliable exclusion of clinically significant pathology. These findings demonstrate that patients can independently acquire clinically interpretable OCT images and that SightSync provides safe, conservative triage performance-supporting its potential as a scalable community-based retinal imaging solution-while a review of unsuccessful scans has identified prototype modifications expected to further improve device feasibility.

Authors

Institutions

Publication Details

Journal
Bioengineering
Published
2026-01-05
DOI
https://doi.org/10.3390/bioengineering13010061
Primary Topic
Optical Coherence Tomography Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Expanding Access to Retinal Imaging Through Patient-Operated Optical Coherence Tomography in a Veterans Affairs Retina Clinic

Warren M. Sobol, Colin K. Drummond, Alan B. Dogan, Katherine G. Barber et al.
Bioengineering
Optical Coherence Tomography Applications
article

Expanding Access to Retinal Imaging Through Patient-Operated Optical Coherence Tomography in a Veterans Affairs Retina Clinic

Warren M. Sobol, Colin K. Drummond, Alan B. Dogan, Katherine G. Barber, Blanche L. Kuo, Brigid C. Devine, Eric S. Eleff, Ankur A. Mehra
article en

Abstract

This study evaluated the feasibility, image quality, and referral accuracy of a patient-operated optical coherence tomography (OCT) device (SightSync) compared with technician-acquired Heidelberg OCT. This study was conducted in a Veterans Affairs retina clinic (Cleveland, Ohio), resulting in a predominantly male (98%) study population representative of the local veteran demographics. One hundred patients attempted self-administered OCT imaging after brief instruction, yielding 118 successful scans (59% of eyes) with no significant association between scan success and age, visual acuity, or diagnosis. Quantitative analysis of 142 paired images showed that SightSync produced interpretable scans with comparable sharpness to Heidelberg OCT, though signal- and intensity-based metrics (signal-to-noise ratio; SNR, contrast-to-noise ratio; CNR, entropy, pixel intensity; p90) were lower, consistent with hardware differences between a compact patient-operated prototype and a clinical-grade system. Among 121 high-quality SightSync scans, referral decisions demonstrated strong agreement with Heidelberg OCT, with a sensitivity of 83.9%, specificity of 75.6%, and a negative predictive value of 93.2%, indicating reliable exclusion of clinically significant pathology. These findings demonstrate that patients can independently acquire clinically interpretable OCT images and that SightSync provides safe, conservative triage performance-supporting its potential as a scalable community-based retinal imaging solution-while a review of unsuccessful scans has identified prototype modifications expected to further improve device feasibility.

BioengineeringVol. 13(1)
University Hospitals of Cleveland (US), Louis Stokes Cleveland VA Medical Center (US), University School (US), Carilion Clinic (US), Case Western Reserve University (US), Virginia Tech (US), Northeast Ohio Medical University (US)
Reduced inequalities
Openalex Percentile: Top 97%
Optical Coherence Tomography Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.