Assessing the feasibility of quantitative SPECT imaging for low 212Pb activity concentrations using anthropomorphic phantoms
We found that larger volumes, such as liver and kidneys with at least 210 Bq/ml, may be quantifiable with an accuracy of approx. 30-40%. While very low activity concentrations below 54 Bq/ml were not quantifiable, this still indicate carefully used imaging results to be of value in dosimetric calculations, also when characterising latter parts of time activity curves.