Interchangeability, accuracy, and precision of a handheld blood lactate analyzer across low to high exercise intensities

Abstract Purpose To assess: (1) within-brand differences and interchangeability of the TaiDoc TD-4289 analyzer, by evaluating the measured blood lactate concentrations across ten units of this analyzer; (2) the accuracy and precision of the analyzer using newly developed strips that present lactate concentrations on a different scale. Methods Venous blood samples were taken at relevant exercise intensities. The blood lactate concentration was measured simultaneously with different units of TaiDoc TD-4289. In the first test, 400 measurements of blood lactate using original strips of code 546 were performed (four intensities times 10 measurement series times 10 units). The within-brand analysis used a modified mean excluding the minimum, maximum, and considered value in each measurement series as reference. In the second test, 64 measurements of blood lactate using newly developed strips of code 335 were performed (eight measurements series times eight units). The measurements were compared to a reference analyzer (YSI Sport 1500) to determine the accuracy and precision when using the code 335 strips. Results Within-brand differences were non-significant; all 10 units were within an absolute mean relative difference to the modified mean of 1.8%. When using code 335 strips, the mean relative difference to the reference analyzer was 45% and the residual standard error was 0.097 mM. Conclusion The TaiDoc TD-4289 analyzer demonstrated high degree of within-brand interchangeability. Using the new code 335 strips, its accuracy was comparable to other established analyzers, while the precision was considerably better than what we have previously found with other commonly used handheld lactate analyzers.

Authors

Institutions

Publication Details

Journal
Sport Sciences for Health
Published
2026-01-05
DOI
https://doi.org/10.1007/s11332-025-01594-5
Primary Topic
Cardiovascular and exercise physiology
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Interchangeability, accuracy, and precision of a handheld blood lactate analyzer across low to high exercise intensities

Fredrik Mentzoni, Thomas Losnegard, Stian Roterud, Martin Skaugen et al.
Sport Sciences for Health
Cardiovascular and exercise physiology
article

Interchangeability, accuracy, and precision of a handheld blood lactate analyzer across low to high exercise intensities

Fredrik Mentzoni, Thomas Losnegard, Stian Roterud, Martin Skaugen, Espen Spro Johansen
article en

Abstract

Abstract Purpose To assess: (1) within-brand differences and interchangeability of the TaiDoc TD-4289 analyzer, by evaluating the measured blood lactate concentrations across ten units of this analyzer; (2) the accuracy and precision of the analyzer using newly developed strips that present lactate concentrations on a different scale. Methods Venous blood samples were taken at relevant exercise intensities. The blood lactate concentration was measured simultaneously with different units of TaiDoc TD-4289. In the first test, 400 measurements of blood lactate using original strips of code 546 were performed (four intensities times 10 measurement series times 10 units). The within-brand analysis used a modified mean excluding the minimum, maximum, and considered value in each measurement series as reference. In the second test, 64 measurements of blood lactate using newly developed strips of code 335 were performed (eight measurements series times eight units). The measurements were compared to a reference analyzer (YSI Sport 1500) to determine the accuracy and precision when using the code 335 strips. Results Within-brand differences were non-significant; all 10 units were within an absolute mean relative difference to the modified mean of 1.8%. When using code 335 strips, the mean relative difference to the reference analyzer was 45% and the residual standard error was 0.097 mM. Conclusion The TaiDoc TD-4289 analyzer demonstrated high degree of within-brand interchangeability. Using the new code 335 strips, its accuracy was comparable to other established analyzers, while the precision was considerably better than what we have previously found with other commonly used handheld lactate analyzers.

Sport Sciences for HealthVol. 22(1)
Norwegian School of Sport Sciences (NO)
Openalex Percentile: Top 97%
Cardiovascular and exercise physiology
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.