Is the tip-apex distance a reliable predictor of implant failure in surgical fixation of femoral neck fractures with the synthes femoral neck system device?

Mean age was 69.6 years old. Mean follow-up was 20.4 months. Median operation duration was 53.5 minutes. Median length of stay was 9 days. Failure rate was 15.2% (n = 7), consisting of two cut-outs, three AVN, one neck shortening and one distal screw breakage. Re-operation rate was 8.7% (n = 4). Mean TAD was 16.8 mm and 18.1 mm in the failure and non-failure groups respectively (P = 0.431) and was not a significant risk factor for failure on logistic regression testing (p = 0.444). Osteoporotic BMD value was more prevalent in the failure group (71.4%) but was not a significant risk factor for failure on logistic regression (P = 0.056). Quality of reduction, age, body mass index (BMI) and pre-operative Garden's score were also insignificant on logistic regression. (P > 0.05) CONCLUSION: The TAD is not a reliable predictor of FNS implant failure for treatment of FNFs. Overall, the FNS is a reliable device with a low failure rate, a short operative time, and a short length of stay comparable to that reported in the literature and with traditional implants.

Authors

Institutions

Publication Details

Journal
BMC Musculoskeletal Disorders
Published
2026-01-02
DOI
https://doi.org/10.1186/s12891-025-09429-7
Primary Topic
Hip and Femur Fractures
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Is the tip-apex distance a reliable predictor of implant failure in surgical fixation of femoral neck fractures with the synthes femoral neck system device?

Christian Hwee Yee Heng, Teddy Cheong, Merng Koon Wong
BMC Musculoskeletal Disorders
Hip and Femur Fractures
article

Is the tip-apex distance a reliable predictor of implant failure in surgical fixation of femoral neck fractures with the synthes femoral neck system device?

Christian Hwee Yee Heng, Teddy Cheong, Merng Koon Wong
article en

Abstract

Mean age was 69.6 years old. Mean follow-up was 20.4 months. Median operation duration was 53.5 minutes. Median length of stay was 9 days. Failure rate was 15.2% (n = 7), consisting of two cut-outs, three AVN, one neck shortening and one distal screw breakage. Re-operation rate was 8.7% (n = 4). Mean TAD was 16.8 mm and 18.1 mm in the failure and non-failure groups respectively (P = 0.431) and was not a significant risk factor for failure on logistic regression testing (p = 0.444). Osteoporotic BMD value was more prevalent in the failure group (71.4%) but was not a significant risk factor for failure on logistic regression (P = 0.056). Quality of reduction, age, body mass index (BMI) and pre-operative Garden's score were also insignificant on logistic regression. (P > 0.05) CONCLUSION: The TAD is not a reliable predictor of FNS implant failure for treatment of FNFs. Overall, the FNS is a reliable device with a low failure rate, a short operative time, and a short length of stay comparable to that reported in the literature and with traditional implants.

BMC Musculoskeletal DisordersVol. 27(1)
Singapore General Hospital (SG), Sengkang General Hospital (SG)
Openalex Percentile: Top 98%
Hip and Femur Fractures
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Is the tip-apex distance a reliable predictor of implant failure in surgical fixation of femoral neck fractures with the synthes femoral neck system device? — Christian Hwee Yee Heng, Teddy Cheong, et al. · BMC Musculoskeletal Disorders (2026) | TGRS Research Map | TGRS