Morphological factors and device deformation associated with recurrence in intracranial aneurysms treated with the contour neurovascular system
Larger neck diameter and Contour deformation are significantly associated with recurrence after treatment with Contour irrespective of the dome-neck ratio. Assessment of Contour deformation on 3D follow-up imaging seems advisable.
Authors
- Fritz Wodarg (ORCID: https://orcid.org/0000-0003-1413-2699)
- Friederike Gärtner (ORCID: https://orcid.org/0000-0001-5957-877X)
- Sönke Peters (ORCID: https://orcid.org/0000-0001-5151-5838)
- Naomi Jean Larsen
- Karim Mostafa (ORCID: https://orcid.org/0000-0002-9372-9006)
- Johannes Hensler (ORCID: https://orcid.org/0000-0002-6594-5334)
- Mariya Pravdivtseva (ORCID: https://orcid.org/0000-0001-9442-088X)
- Sylvia Saalfeld
Institutions
- University Hospital Schleswig-Holstein (DE)
- University of Lübeck (DE)
Publication Details
- Journal
- Journal of Neuroradiology
- Published
- 2026-01-02
- DOI
- https://doi.org/10.1016/j.neurad.2025.101407
- Primary Topic
- Intracranial Aneurysms: Treatment and Complications
- Type
- article
- Field-Weighted Citation Impact
- 0.00