Evaluation of the Accuracy of Output Values of Torque Wrenches in Clinical Settings

Aim: The aim of this in vitro study was to determine and compare the accuracy of different types of mechanical torque limiting devices currently used in clinical settings. Materials and Methods: A total of 30 torque wrenches including both friction-style (n=17) and springstyle (n=13) instruments were evaluated in this study. A calibrated digital torque meter (Lutron TQ-8800; Lutron Electronic Enterprise Co., Ltd.; Taiwan) was used to measure the output torque values of the torque wrenches. Measurements were repeated 10 times for each wrench and reported in Ncm. The Shapiro-Wilk test was performed for mean torque measurment values followed by an independent t-test and the percentages of absolute deviation from target torque values were calculated. The results were evaluated at a 95% confidence interval with significance at p<0.05. Results: The mean torque measurement values of friction-style devices showed a statistically significant difference from target torque values (p<0.05), while the mean torque measurement values of spring-style devices did not differ significantly from target values (p>0.05). While the percentage deviation values ranged between 7.64 and 11.76 for friction-style devices, they were 1.79- 5.12 for spring-style devices. Conclusion: The accuracy of spring-type torque wrenches is better than that of friction-type torque wrenches, and the calibration of all wrenches should routinely be performed at the target torque value specified by the manufacturer to ensure optimum healthcare services.

Authors

Institutions

Publication Details

Journal
Necmettin Erbakan Universitesi Dis Hekimligi Dergisi (NEU Dent J) Necmettin Erbakan
Published
2025-12-30
DOI
https://doi.org/10.51122/neudentj.2025.169
Primary Topic
Musculoskeletal pain and rehabilitation
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of the Accuracy of Output Values of Torque Wrenches in Clinical Settings

Mert Tayşi, Ayşegül ERTEN TAYŞİ, Pınar ERCAL
Necmettin Erbakan Universitesi Dis Hekimligi Dergisi (NEU Dent J) Necmettin Erbakan
Musculoskeletal pain and rehabilitation
article

Evaluation of the Accuracy of Output Values of Torque Wrenches in Clinical Settings

Mert Tayşi, Ayşegül ERTEN TAYŞİ, Pınar ERCAL
article en

Abstract

Aim: The aim of this in vitro study was to determine and compare the accuracy of different types of mechanical torque limiting devices currently used in clinical settings. Materials and Methods: A total of 30 torque wrenches including both friction-style (n=17) and springstyle (n=13) instruments were evaluated in this study. A calibrated digital torque meter (Lutron TQ-8800; Lutron Electronic Enterprise Co., Ltd.; Taiwan) was used to measure the output torque values of the torque wrenches. Measurements were repeated 10 times for each wrench and reported in Ncm. The Shapiro-Wilk test was performed for mean torque measurment values followed by an independent t-test and the percentages of absolute deviation from target torque values were calculated. The results were evaluated at a 95% confidence interval with significance at p<0.05. Results: The mean torque measurement values of friction-style devices showed a statistically significant difference from target torque values (p<0.05), while the mean torque measurement values of spring-style devices did not differ significantly from target values (p>0.05). While the percentage deviation values ranged between 7.64 and 11.76 for friction-style devices, they were 1.79- 5.12 for spring-style devices. Conclusion: The accuracy of spring-type torque wrenches is better than that of friction-type torque wrenches, and the calibration of all wrenches should routinely be performed at the target torque value specified by the manufacturer to ensure optimum healthcare services.

Necmettin Erbakan Universitesi Dis Hekimligi Dergisi (NEU Dent J) Necmettin Erbakan
Istanbul University-Cerrahpaşa (TR)
Openalex Percentile: Top 29%
Musculoskeletal pain and rehabilitation
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.