Analytical and diagnostic evaluation of the Anvajo fluidlab 2 analyzer: a novel urine particle analyzer for clinical application using digital holographic microscopy?
Fluidlab 2 provides desirable imprecision for RBC and WBC, and meets the criteria for linearity and LoQ. Cohen's weighted kappa coefficients show an optimal comparison to visual microscopy for RBC, WBC and SEC and a minimum comparison for casts and non-SEC. This evaluation demonstrated promising results for the use of the fluidlab 2 analyzer in a clinical setting to detect kidney-related diseases based on urine particle analysis.