Agreement between arterial blood pressures measured non-invasively and invasively in anaesthetised sheep.

Background Both invasive and non-invasive blood pressure measuring methods are used in clinical and experimental work in veterinary medicine and several studies validating these methods are published. The aim of this study was to assess the level of agreement between non-invasive and invasive arterial blood pressure measurements within a wide range of blood pressures in sheep. Six adult Texel-cross female sheep were included. Anaesthesia was induced with IV propofol (4-8 mg/kg) and maintained with sevoflurane in 50% oxygen and air. Blood pressure measurements were simultaneously obtained using both invasive methods (electronic pressure transducer (EPT) connected to the auricular and carotid arteries, and aneroid manometer connected to the auricular artery) and non-invasive methods (the oscillometric device (OD) and Doppler). Carotid artery EPT was considered the 'gold standard' to which other methods were compared. The agreement between the two methods was evaluated with the Bland-Altman method, in conditions of normotension, hypertension (MAP > 160 mmHg, induced with phenylephrine), and hypotension (MAP < 50 mmHg, induced with acepromazine). The devices were evaluated using the American College of Veterinary Internal Medicine guidelines.Results Carotid MAP values ranged from 37 to 192 mmHg. Mean bias and limits of agreement were - 2.8 mmHg and - 11-5.4 mmHg for the auricular EPT MAP; 5.1 mmHg and - 12.2-22.3 mmHg for the aneroid manometer MAP; - 2.7 mmHg and - 23.9-18.6 mmHg for the OD MAP; - 1.5 mmHg and - 20.1-23.2 mmHg for the Doppler SAP; respectively. Correlation coefficients for all methods were > 0.95.Conclusions Doppler SAP and OD MAP demonstrated acceptable accuracy over a broad blood pressure range, supporting their utility for sheep as surrogate of invasive methods of measuring blood pressure. Auricular EPT was the best surrogate of central arterial pressure to be used in clinical conditions.

Authors

Institutions

Publication Details

Journal
Open Access CRIS of the University of Bern
Published
2025-12-08
DOI
https://doi.org/10.48620/93453
Primary Topic
Veterinary Pharmacology and Anesthesia
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Agreement between arterial blood pressures measured non-invasively and invasively in anaesthetised sheep.

Magdy Adam, Sanna Sainmaa, Anna Mykkänen, Anna Valldeoriola Cardó et al.
Open Access CRIS of the University of Bern
Veterinary Pharmacology and Anesthesia
article

Agreement between arterial blood pressures measured non-invasively and invasively in anaesthetised sheep.

Magdy Adam, Sanna Sainmaa, Anna Mykkänen, Anna Valldeoriola Cardó, Marja Riitta Raekallio, Anna Vilhelmiina Huuskonen, Daniela C. Casoni
article en

Abstract

Background Both invasive and non-invasive blood pressure measuring methods are used in clinical and experimental work in veterinary medicine and several studies validating these methods are published. The aim of this study was to assess the level of agreement between non-invasive and invasive arterial blood pressure measurements within a wide range of blood pressures in sheep. Six adult Texel-cross female sheep were included. Anaesthesia was induced with IV propofol (4-8 mg/kg) and maintained with sevoflurane in 50% oxygen and air. Blood pressure measurements were simultaneously obtained using both invasive methods (electronic pressure transducer (EPT) connected to the auricular and carotid arteries, and aneroid manometer connected to the auricular artery) and non-invasive methods (the oscillometric device (OD) and Doppler). Carotid artery EPT was considered the 'gold standard' to which other methods were compared. The agreement between the two methods was evaluated with the Bland-Altman method, in conditions of normotension, hypertension (MAP > 160 mmHg, induced with phenylephrine), and hypotension (MAP < 50 mmHg, induced with acepromazine). The devices were evaluated using the American College of Veterinary Internal Medicine guidelines.Results Carotid MAP values ranged from 37 to 192 mmHg. Mean bias and limits of agreement were - 2.8 mmHg and - 11-5.4 mmHg for the auricular EPT MAP; 5.1 mmHg and - 12.2-22.3 mmHg for the aneroid manometer MAP; - 2.7 mmHg and - 23.9-18.6 mmHg for the OD MAP; - 1.5 mmHg and - 20.1-23.2 mmHg for the Doppler SAP; respectively. Correlation coefficients for all methods were > 0.95.Conclusions Doppler SAP and OD MAP demonstrated acceptable accuracy over a broad blood pressure range, supporting their utility for sheep as surrogate of invasive methods of measuring blood pressure. Auricular EPT was the best surrogate of central arterial pressure to be used in clinical conditions.

Open Access CRIS of the University of Bern
Central Institute for Experimental Animals (JP), Institute of Experimental Physics of the Slovak Academy of Sciences (SK)
Good health and well-being
Openalex Percentile: Top 25%
Veterinary Pharmacology and Anesthesia
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.