Biomechanics, Material Performance, and Wear Analysis in Total Hip Arthroplasty: A Review

Total hip arthroplasty (THA) is a transformative procedure for managing severe hip disorders, yet implant longevity remains a critical challenge, particularly for younger, active patients. Wear-related complications are a leading cause of revision surgery, emphasizing the need for optimized design and material performance. This systematic review aims to synthesize evidence on the wear behavior, material properties, and design parameters of hip implants with a focus on finite element analysis (FEA)-based predictive approaches. A comprehensive literature search was conducted across Scopus, PubMed, ScienceDirect, MDPI, and Cochrane databases following PRISMA guidelines. Studies published between 2010 and 2025 were included if they addressed THA biomechanics, wear analysis, or material optimization using FEA, hip simulators, or radiostereometric techniques. Key findings reveal that larger femoral heads, while reducing contact pressure, increase wear due to greater sliding distance. Gravimetric wear rates ranged from 3.15 ± 0.27 mg/Mc to 2.18 ± 0.31 mg/Mc, while linear and volumetric wear reached 0.0375 mm/Mc and 33.6 mm3/Mc, respectively. Stress analysis showed custom stems exhibited markedly lower von Mises stress (194.9 MPa) compared to standard designs (664.3 MPa), and fatigue assessments confirmed a factor of safety > 1 across profiles. Patient-specific factors, such as body weight, significantly influenced wear with a 26% increase in metal wear observed between 100 kg and 140 kg. This systematic review synthesizes current research on total hip arthroplasty, emphasizing biomechanical and material factors critical for implant longevity and patient care. It uniquely integrates FEA-based wear prediction with clinical implications, bridging computational modeling, geometry optimization, and material performance to provide actionable insights for next-generation, patient-specific hip implant design.

Authors

Institutions

Publication Details

Journal
Surgeries
Published
2025-12-30
DOI
https://doi.org/10.3390/surgeries7010007
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Biomechanics, Material Performance, and Wear Analysis in Total Hip Arthroplasty: A Review

Şenay Mihçin, B. Satish Shenoy, Nishant Nikam, Dr Shyamasunder Bhat N et al.
Surgeries
Orthopaedic implants and arthroplasty
article

Biomechanics, Material Performance, and Wear Analysis in Total Hip Arthroplasty: A Review

Şenay Mihçin, B. Satish Shenoy, Nishant Nikam, Dr Shyamasunder Bhat N, Sawan Shetty, Chethan K. N., Laxmikant G. Keni
article en

Abstract

Total hip arthroplasty (THA) is a transformative procedure for managing severe hip disorders, yet implant longevity remains a critical challenge, particularly for younger, active patients. Wear-related complications are a leading cause of revision surgery, emphasizing the need for optimized design and material performance. This systematic review aims to synthesize evidence on the wear behavior, material properties, and design parameters of hip implants with a focus on finite element analysis (FEA)-based predictive approaches. A comprehensive literature search was conducted across Scopus, PubMed, ScienceDirect, MDPI, and Cochrane databases following PRISMA guidelines. Studies published between 2010 and 2025 were included if they addressed THA biomechanics, wear analysis, or material optimization using FEA, hip simulators, or radiostereometric techniques. Key findings reveal that larger femoral heads, while reducing contact pressure, increase wear due to greater sliding distance. Gravimetric wear rates ranged from 3.15 ± 0.27 mg/Mc to 2.18 ± 0.31 mg/Mc, while linear and volumetric wear reached 0.0375 mm/Mc and 33.6 mm3/Mc, respectively. Stress analysis showed custom stems exhibited markedly lower von Mises stress (194.9 MPa) compared to standard designs (664.3 MPa), and fatigue assessments confirmed a factor of safety > 1 across profiles. Patient-specific factors, such as body weight, significantly influenced wear with a 26% increase in metal wear observed between 100 kg and 140 kg. This systematic review synthesizes current research on total hip arthroplasty, emphasizing biomechanical and material factors critical for implant longevity and patient care. It uniquely integrates FEA-based wear prediction with clinical implications, bridging computational modeling, geometry optimization, and material performance to provide actionable insights for next-generation, patient-specific hip implant design.

SurgeriesVol. 7(1)
Izmir Institute of Technology (TR), Manipal Academy of Higher Education (IN), Kasturba Medical College, Manipal (IN)
Openalex Percentile: Top 25%
Orthopaedic implants and arthroplasty
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.