Environmental monitoring of e.g. radionuclides are important in order to establish a baseline of radionuclides present in the environment, e.g. in order to detect increases in activites. Often low activities are present which require long measurement times. This work presents improvements with respect to reducing instrument time required when measuring to a pre-determined MDA (Minimum Detectable Activity). The method studied, involved the separation of <sup>137</sup>Cs in tap water samples using AMP-PAN (ammonium molybdophosphate in polyacetonitrile) combined with IDGS (isotope dilution gamma spectrometry). The results show that significantly lower instrument times can be used when measuring to a pre-determined MDA (considered fit-for-purpose), if the same amount of sample can be measured in a more efficient geometry. Moreover, using more sample volume will also decrease the instrument time significantly. Measurement times will, for increased efficiencies, approximately be shortened by a factor equal to the square of the ratio between the efficiencies. Likewise, instrument time will approximately be reduced by a factor equal to the square of the ratios of sample volumes when sample volume is increased. The same relationship is valid for chemical yield, emphasising the importance of high chemical yields. When implementing AMP-separation of <sup>137</sup>Cs and doubling the sample volume compared to the original laboratory method (evaporation), the instrument time needed to reach the target MDA (0.60 mBq/L) was reduced by a factor of about eight. Apart from a much shorter measurement time, the time for sample preparation was shortened from a few days to about 2 h.

Authors

Institutions

Publication Details

Journal
Applied Radiation and Isotopes
Published
2025-12-31
DOI
https://doi.org/10.1016/j.apradiso.2025.112413
Primary Topic
Radioactive contamination and transfer
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The effect of increased sample volume and peak efficiency on instrument time when measuring to a pre-determined MDA: Applied to AMP-separation of 137Cs in tap-water samples

H. Ramebäck, M. Östman, K. Norlin
Applied Radiation and Isotopes
Radioactive contamination and transfer
article

The effect of increased sample volume and peak efficiency on instrument time when measuring to a pre-determined MDA: Applied to AMP-separation of 137Cs in tap-water samples

H. Ramebäck, M. Östman, K. Norlin
article en

Abstract

Environmental monitoring of e.g. radionuclides are important in order to establish a baseline of radionuclides present in the environment, e.g. in order to detect increases in activites. Often low activities are present which require long measurement times. This work presents improvements with respect to reducing instrument time required when measuring to a pre-determined MDA (Minimum Detectable Activity). The method studied, involved the separation of <sup>137</sup>Cs in tap water samples using AMP-PAN (ammonium molybdophosphate in polyacetonitrile) combined with IDGS (isotope dilution gamma spectrometry). The results show that significantly lower instrument times can be used when measuring to a pre-determined MDA (considered fit-for-purpose), if the same amount of sample can be measured in a more efficient geometry. Moreover, using more sample volume will also decrease the instrument time significantly. Measurement times will, for increased efficiencies, approximately be shortened by a factor equal to the square of the ratio between the efficiencies. Likewise, instrument time will approximately be reduced by a factor equal to the square of the ratios of sample volumes when sample volume is increased. The same relationship is valid for chemical yield, emphasising the importance of high chemical yields. When implementing AMP-separation of <sup>137</sup>Cs and doubling the sample volume compared to the original laboratory method (evaporation), the instrument time needed to reach the target MDA (0.60 mBq/L) was reduced by a factor of about eight. Apart from a much shorter measurement time, the time for sample preparation was shortened from a few days to about 2 h.

Applied Radiation and IsotopesVol. 229
Swedish Defence Research Agency (SE)
Clean water and sanitation
Openalex Percentile: Top 35%
Radioactive contamination and transfer
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.