Challenges in accurate HDV RNA quantification: inter-assay variability and the impact of thermal shock

Quantitative RT-PCR (qRT-PCR) is essential for monitoring hepatitis delta virus (HDV) RNA, yet assays lack standardization. We aimed to evaluate the performance of three qRT-PCR assays and to assess the impact of a pre-analytical thermal shock procedure. We conducted a comparative study using 206 samples (106 with anti-HDV antibodies and 100 anti-HDV-negative as a control group), which were tested in parallel with three qRT-PCR assays: Vircell, Certest, and Altona. Performance was evaluated for inter-assay agreement (kappa index), quantitative correlation (R²), and bias (Bland-Altman). Altona detected 56 HDV-RNA positive samples, whereas Vircell and Certest detected 55 positive samples. Inter-assay agreement was perfect comparing Vircell vs Certest (agreement = 100%, к = 1.000) and almost perfect comparing Altona with both Vircell and Certest (agreement = 99.5%, к = 0.988). Quantitatively, Vircell and Certest assays showed relevant systematic biases compared to Altona, overestimating viral loads by approximately 0.24 log<sub>10</sub> International Units (IU)/mL (Certest) and 0.33 log<sub>10</sub> IU/mL (Vircell). The correlation with Altona was strong for Certest (R² = 0.864) and moderate for Vircell (R² = 0.793), while the correlation between Certest and Vircell was weaker (R² = 0.720). Thermal shock improved sensitivity in one case (Certest vs Vircell) but increased quantitative variability, worsening the inter-assay correlation (R² = 0.684). In conclusion, all three assays were highly concordant for the qualitative diagnosis of HDV infection, but their quantitative biases prevent their interchangeable use for treatment monitoring. Thermal shock is not recommended for routine monitoring, as a significant compromise in quantitative accuracy and precision outweighs any potential gains in sensitivity.IMPORTANCEThis study evaluates three hepatitis delta virus (HDV) RNA quantitative real-time PCR (qRT-PCR) assays, crucial for managing the HDV infection, particularly in the setting of new therapies like Bulevirtide, where assessing viral load reduction and accurate monitoring is paramount. We reveal significant quantitative biases among widely used assays, precluding their interchangeable use and risking misinterpretation of treatment response. Furthermore, our systematic assessment of the thermal shock pre-analytical procedure highlights its detrimental impact on quantitative precision, despite modest sensitivity gains. This work provides essential evidence for clinicians and laboratories, guiding assay selection and standardization efforts to optimize HDV diagnosis and patient monitoring.

Authors

Institutions

Publication Details

Journal
Journal of Clinical Microbiology
Published
2025-12-31
DOI
https://doi.org/10.1128/jcm.01517-25
Primary Topic
Hepatitis B Virus Studies
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Challenges in accurate HDV RNA quantification: inter-assay variability and the impact of thermal shock

Luís Miguel Real, Felipe Pérez‐García, Isidoro Martínez, Mario Muñoz Monte et al.
Journal of Clinical Microbiology
Hepatitis B Virus Studies
article

Challenges in accurate HDV RNA quantification: inter-assay variability and the impact of thermal shock

Luís Miguel Real, Felipe Pérez‐García, Isidoro Martínez, Mario Muñoz Monte, Sandra Rodrigo, Carlos Pita-Martínez, Ana Virseda-Berdices, Helena Codina, Juan Macías, R. Muley Alonso, Lara Mesones, Juan Cuadros-González, Salvador Resino, Daniel Sepúlveda-Crespo
article en

Abstract

Quantitative RT-PCR (qRT-PCR) is essential for monitoring hepatitis delta virus (HDV) RNA, yet assays lack standardization. We aimed to evaluate the performance of three qRT-PCR assays and to assess the impact of a pre-analytical thermal shock procedure. We conducted a comparative study using 206 samples (106 with anti-HDV antibodies and 100 anti-HDV-negative as a control group), which were tested in parallel with three qRT-PCR assays: Vircell, Certest, and Altona. Performance was evaluated for inter-assay agreement (kappa index), quantitative correlation (R²), and bias (Bland-Altman). Altona detected 56 HDV-RNA positive samples, whereas Vircell and Certest detected 55 positive samples. Inter-assay agreement was perfect comparing Vircell vs Certest (agreement = 100%, к = 1.000) and almost perfect comparing Altona with both Vircell and Certest (agreement = 99.5%, к = 0.988). Quantitatively, Vircell and Certest assays showed relevant systematic biases compared to Altona, overestimating viral loads by approximately 0.24 log<sub>10</sub> International Units (IU)/mL (Certest) and 0.33 log<sub>10</sub> IU/mL (Vircell). The correlation with Altona was strong for Certest (R² = 0.864) and moderate for Vircell (R² = 0.793), while the correlation between Certest and Vircell was weaker (R² = 0.720). Thermal shock improved sensitivity in one case (Certest vs Vircell) but increased quantitative variability, worsening the inter-assay correlation (R² = 0.684). In conclusion, all three assays were highly concordant for the qualitative diagnosis of HDV infection, but their quantitative biases prevent their interchangeable use for treatment monitoring. Thermal shock is not recommended for routine monitoring, as a significant compromise in quantitative accuracy and precision outweighs any potential gains in sensitivity.IMPORTANCEThis study evaluates three hepatitis delta virus (HDV) RNA quantitative real-time PCR (qRT-PCR) assays, crucial for managing the HDV infection, particularly in the setting of new therapies like Bulevirtide, where assessing viral load reduction and accurate monitoring is paramount. We reveal significant quantitative biases among widely used assays, precluding their interchangeable use and risking misinterpretation of treatment response. Furthermore, our systematic assessment of the thermal shock pre-analytical procedure highlights its detrimental impact on quantitative precision, despite modest sensitivity gains. This work provides essential evidence for clinicians and laboratories, guiding assay selection and standardization efforts to optimize HDV diagnosis and patient monitoring.

Journal of Clinical MicrobiologyVol. 64(2)
Universidad Complutense de Madrid (ES), Universidad de Alcalá (ES), Instituto de Salud Carlos III (ES), Hospital General Universitario Gregorio Marañón (ES), Centro de Investigación Biomédica en Red de Enfermedades Respiratorias (ES), Centro Nacional de Microbiologia (ES), Hospital Universitario Príncipe de Asturias (ES), Centro de Investigación Biomédica en Red (ES), Instituto de Biomedicina de Sevilla (ES), Hospital Universitario de Valme (ES), Universidad de Sevilla (ES)
European Society of Clinical Microbiology and Infectious Diseases, Ministerio de Ciencia e Innovación, Instituto de Salud Carlos III
Good health and well-being
Openalex Percentile: Top 29%
Hepatitis B Virus Studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.