Numerical study of tibial inserts made of conventional and vitamin E blended UHMWPE with and without cross linking in total knee replacement under EHL conditions

Total Knee Replacement (TKR) effectively improves mobility and reduces pain in patients with severe arthritis. Polyethylene wear limits implant longevity by inducing osteolysis and aseptic loosening. Therefore, this study evaluates the lubrication performance of four clinically relevant UHMWPE formulations representing distinct combination of crosslinking dosage and antioxidant. An ellipsoidal on-plane model of an artificial knee joint is considered and all governing equations are solved using Newton Raphson method. The results indicate that, vitamin E blended UHMWPE without cross linking exhibits highest contact pressure with only a marginal reduction in film thickness, while achieving the lowest coefficient of friction under EHL conditions.

Authors

Institutions

Publication Details

Journal
Computer Methods in Biomechanics & Biomedical Engineering
Published
2025-12-30
DOI
https://doi.org/10.1080/10255842.2025.2610678
Primary Topic
Total Knee Arthroplasty Outcomes
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Numerical study of tibial inserts made of conventional and vitamin E blended UHMWPE with and without cross linking in total knee replacement under EHL conditions

P. Bajpai, Rekha Bali
Computer Methods in Biomechanics & Biomedical Engineering
Total Knee Arthroplasty Outcomes
article

Numerical study of tibial inserts made of conventional and vitamin E blended UHMWPE with and without cross linking in total knee replacement under EHL conditions

P. Bajpai, Rekha Bali
article en

Abstract

Total Knee Replacement (TKR) effectively improves mobility and reduces pain in patients with severe arthritis. Polyethylene wear limits implant longevity by inducing osteolysis and aseptic loosening. Therefore, this study evaluates the lubrication performance of four clinically relevant UHMWPE formulations representing distinct combination of crosslinking dosage and antioxidant. An ellipsoidal on-plane model of an artificial knee joint is considered and all governing equations are solved using Newton Raphson method. The results indicate that, vitamin E blended UHMWPE without cross linking exhibits highest contact pressure with only a marginal reduction in film thickness, while achieving the lowest coefficient of friction under EHL conditions.

Computer Methods in Biomechanics & Biomedical Engineering
Harcourt Butler Technical University (IN)
Openalex Percentile: Top 25%
Total Knee Arthroplasty Outcomes
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Numerical study of tibial inserts made of conventional and vitamin E blended UHMWPE with and without cross linking in total knee replacement under EHL conditions — P. Bajpai, Rekha Bali · Computer Methods in Biomechanics & Biomedical Engineering (2025) | TGRS Research Map | TGRS