Comparison of blood pressure measurements between the Huawei Watch D smartwatch application and the validated Omron M3 Intellisense device: Observational study

Popular wearable devices that record a variety of health metrics, such as real-time blood pressure (BP), could play a role in detecting hypertension in the population. The objective of this study was to compare the validity of the Huawei D device with that of the Omron M3 Intellisense device used by health workers. Sampling was used to obtain patients with normal and abnormal BP. Measurements were taken on the wrist via the Huawei Watch D, while measurements were made using the Omron M3 Intellisense (prevalidated) device used by experienced healthcare personnel at the health centre. A total of 100 normotensive and 100 hypertensive patients were analysed. The concordance statistics were calculated, and linear regression analysis was performed. The intraclass correlation coefficient was calculated to assess the concordance between the instruments used in the measurement and among the observers who made the measurements under a two-factor mixed-effects model, with calculation of the degree of absolute concordance in mean measurements. We found moderate results (systolic blood pressure (SBP) at 30 and 60 min and heart rate at 30 min) and nil for the rest [0.63 (0.36-0.78) < 0.001]; in addition, the comparison of the two groups was verified by a Bland-Altamn graph, in which the disparity of the results was observed. The Huawei D BP measuring device was shown to be relatively effective for measuring BP in hypertensive patients. However, the specificity of BP measurement in normotensive patients was not as reliable as in hypertensive patients.

Authors

Institutions

Publication Details

Journal
Experimental Physiology
Published
2025-12-26
DOI
https://doi.org/10.1113/ep093286
Primary Topic
Blood Pressure and Hypertension Studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparison of blood pressure measurements between the Huawei Watch D smartwatch application and the validated Omron M3 Intellisense device: Observational study

Raúl Frutos Llanes, David Rodríguez‐Sanz, Angélica Campón Chekroun, José Manuel Barragán Casas et al.
Experimental Physiology
Blood Pressure and Hypertension Studies
article

Comparison of blood pressure measurements between the Huawei Watch D smartwatch application and the validated Omron M3 Intellisense device: Observational study

Raúl Frutos Llanes, David Rodríguez‐Sanz, Angélica Campón Chekroun, José Manuel Barragán Casas, Rosario Pastor Martín, Zacarías Sánchez Milá, Sonia Jiménez Blanco, Jorge Velázquez Saornil
article en

Abstract

Popular wearable devices that record a variety of health metrics, such as real-time blood pressure (BP), could play a role in detecting hypertension in the population. The objective of this study was to compare the validity of the Huawei D device with that of the Omron M3 Intellisense device used by health workers. Sampling was used to obtain patients with normal and abnormal BP. Measurements were taken on the wrist via the Huawei Watch D, while measurements were made using the Omron M3 Intellisense (prevalidated) device used by experienced healthcare personnel at the health centre. A total of 100 normotensive and 100 hypertensive patients were analysed. The concordance statistics were calculated, and linear regression analysis was performed. The intraclass correlation coefficient was calculated to assess the concordance between the instruments used in the measurement and among the observers who made the measurements under a two-factor mixed-effects model, with calculation of the degree of absolute concordance in mean measurements. We found moderate results (systolic blood pressure (SBP) at 30 and 60 min and heart rate at 30 min) and nil for the rest [0.63 (0.36-0.78) < 0.001]; in addition, the comparison of the two groups was verified by a Bland-Altamn graph, in which the disparity of the results was observed. The Huawei D BP measuring device was shown to be relatively effective for measuring BP in hypertensive patients. However, the specificity of BP measurement in normotensive patients was not as reliable as in hypertensive patients.

Experimental Physiology
Universidad Complutense de Madrid (ES), Catholic University of Ávila (ES), Universidad Pontificia de Salamanca (ES), Centro de Cirugía de Mínima Invasión Jesús Usón (ES)
Openalex Percentile: Top 31%
Blood Pressure and Hypertension Studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.