Integrated 3D-Printed Microfluidic Device for Immunocapture and Electrochemical Assessment of Transferrin Saturation in <i>Point-of-Care</i> Stroke Diagnostics

A 3D-printed electrochemical microfluidic device (3D-EMD) was developed to assess the transferrin saturation (TSAT) biomarker in ischemic stroke patients. The all-in-one 3D-EMD integrates a strategically engineered immunoassay module for the direct and selective isolation of transferrin (Tf) from unpretreated clinical samples, unaffected by sample coloration, with an interchangeable electrochemical sensor for the simultaneous detection of Tf and Tf-bound iron. Both modules are interconnected through microfluidic channels whose flow is regulated by a cylindrical rotary valve. The analytical workflow enables magnetic bead-based direct Tf immunocapture and simultaneous electrochemical detection of Tf and Tf-bound iron via square wave voltammetry, allowing TSAT assessment within 60 min using only 50 μL of sample. Validation with certified reference materials demonstrated excellent accuracy (<i>E</i><sub>r</sub> ≤ 5%) and precision (RSD ≤ 4%). Application to human serum from ischemic stroke patients showed strong correlation (<i>r</i> = 0.87) and agreement (slope 0.9 ± 0.3; intercept 6 ± 10; <i>p</i> < 0.05) with the urea-PAGE reference method, which typically requires up to 18 h. Overall, the 3D-EMD constitutes an elegant, fully integrated dual-functionality platform that seamlessly combines customizable sample preparation with online electrochemical detection in a single device. This configuration enables direct serum analysis and supports clinical decision-making in time-critical conditions. The device shows strong potential as a rapid <i>point-of-care</i> testing candidate for ischemic stroke and as a next-generation platform for broader clinical diagnostics.

Authors

Institutions

Publication Details

Journal
ACS Sensors
Published
2025-12-26
DOI
https://doi.org/10.1021/acssensors.5c02834
Primary Topic
Electrochemical sensors and biosensors
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Integrated 3D-Printed Microfluidic Device for Immunocapture and Electrochemical Assessment of Transferrin Saturation in Point-of-Care Stroke Diagnostics

Silvia Dortéz, Teresa Gasull, Alberto Escarpa, Davide Paolini et al.
ACS Sensors
Electrochemical sensors and biosensors
article

Integrated 3D-Printed Microfluidic Device for Immunocapture and Electrochemical Assessment of Transferrin Saturation in Point-of-Care Stroke Diagnostics

Silvia Dortéz, Teresa Gasull, Alberto Escarpa, Davide Paolini, Marta Pacheco, Flavio Della Pelle, Dario Compagnone
article en

Abstract

A 3D-printed electrochemical microfluidic device (3D-EMD) was developed to assess the transferrin saturation (TSAT) biomarker in ischemic stroke patients. The all-in-one 3D-EMD integrates a strategically engineered immunoassay module for the direct and selective isolation of transferrin (Tf) from unpretreated clinical samples, unaffected by sample coloration, with an interchangeable electrochemical sensor for the simultaneous detection of Tf and Tf-bound iron. Both modules are interconnected through microfluidic channels whose flow is regulated by a cylindrical rotary valve. The analytical workflow enables magnetic bead-based direct Tf immunocapture and simultaneous electrochemical detection of Tf and Tf-bound iron via square wave voltammetry, allowing TSAT assessment within 60 min using only 50 μL of sample. Validation with certified reference materials demonstrated excellent accuracy (<i>E</i><sub>r</sub> ≤ 5%) and precision (RSD ≤ 4%). Application to human serum from ischemic stroke patients showed strong correlation (<i>r</i> = 0.87) and agreement (slope 0.9 ± 0.3; intercept 6 ± 10; <i>p</i> < 0.05) with the urea-PAGE reference method, which typically requires up to 18 h. Overall, the 3D-EMD constitutes an elegant, fully integrated dual-functionality platform that seamlessly combines customizable sample preparation with online electrochemical detection in a single device. This configuration enables direct serum analysis and supports clinical decision-making in time-critical conditions. The device shows strong potential as a rapid <i>point-of-care</i> testing candidate for ischemic stroke and as a next-generation platform for broader clinical diagnostics.

ACS SensorsVol. 11(1)
University of Teramo (IT), Universidad de Alcalá (ES), Deltares (NL), Institut d'Investigació en Ciències de la Salut Germans Trias i Pujol (ES), Cellular Research (United States) (US)
Openalex Percentile: Top 42%
Electrochemical sensors and biosensors
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.