Comparative Dosimetry of Handheld Versus Wall-Mounted Intraoral X-ray Systems: A Phantom-Based Study
Despite comparable diagnostic image quality, the handheld dental X-ray unit delivered substantially higher radiation doses to both the operator and the patient-equivalent phantom than the wall-mounted system. Wall-mounted units should remain the standard where patient mobility permits, reserving handheld systems for clinically justified scenarios.
Authors
- Niti Dharmendra Shah
- Abdul Rafi Shaik
- Sai Priyanka Gaddipati
- Rahul VC Tiwari
- Vignesh Guptha Raju
- Rajsandeep Singh
- Venkat Hemant Akurati
- Seema Gupta
Institutions
- University of Louisville (US)
- RKDF University (IN)
- Kothiwal Dental College and Research Centre (IN)
- Narayana Dental College and Hospital (IN)
- Banner Thunderbird Medical Center (US)
- P.M.N.M. Dental College and Hospital (IN)
- Punjab Engineering College (IN)
- Ragas Dental College & Hospital (IN)
Publication Details
- Journal
- Cureus
- Published
- 2025-12-25
- DOI
- https://doi.org/10.7759/cureus.100091
- Primary Topic
- Dental Radiography and Imaging
- Type
- article
- Field-Weighted Citation Impact
- 0.00