Patient‐specific quality assurance at the Heidelberg Ion Beam Therapy Center: 10 years experience in treatment plan verification
PSQA at HIT demonstrated high dosimetric accuracy and delivery stability. Integration of IDCs and log file analysis may improve efficiency and allow to omit verification measurements in well-established cases without compromising patient safety and treatment quality, if the extensive machine QA program is maintained.
Authors
- K. Henkner
- Abdallah Qubala (ORCID: https://orcid.org/0000-0002-6526-2453)
- Christian P. Karger (ORCID: https://orcid.org/0000-0002-7194-8136)
- M. Ellerbrock
- John Horn
- Oliver Jäkel
- Marcus Winter
Institutions
- German Cancer Research Center (DE)
- Heidelberg University (DE)
- University Hospital Heidelberg (DE)
- Heidelberg University of Education (DE)
- Heidelberger Institut für Radioonkologie (DE)
- National Center for Tumor Diseases (DE)
- Heidelberg Engineering (Germany) (DE)
- Heidelberg University (US)
Publication Details
- Journal
- Medical Physics
- Published
- 2025-12-26
- DOI
- https://doi.org/10.1002/mp.70237
- Primary Topic
- Radiation Therapy and Dosimetry
- Type
- article
- Field-Weighted Citation Impact
- 0.00