Screening capabilities and prevalence of sleep-disordered breathing in patients with implantable cardioverter-defibrillators

Objective. To assess diagnostic efficacy of screening methods for sleep-disordered breathing — questionnaires (assessing the risk of obstructive sleep apnea and daytime sleepiness), overnight pulse oximetry and apnea detection algorithm embedded in implantable cardioverter-defibrillators (ICDs) in patients with ICDs. Material and methods. The study included 67 patients aged 34-83 years (median 65). There were 54 (81%) men, median body mass index was 29.4 kg/m2. All patients underwent CRM to detect sleep-disordered breathing. Screening performance of various diagnostic methods for sleep apnea detection was assessed in 39 patients using STOP-Bang questionnaire and Epworth Sleepiness Scale (ESS), in 20 patients — using computer-based pulse oximetry (PO), in 7 patients — via interpretation of respiratory disturbance index obtained from ICDs of specific models. Results of these screening methods were compared with CRM data. Results. According to CRM data, sleep-disordered breathing was detected in 64 (95.5%) patients (mild in 23 (34.5%), moderate in 22 (33%), and severe in 19 (28%) cases). In all cases, sleep-disordered breathing was obstructive, but 19 (28%) patients had combination with significant (more than 5 events/hour) episodes of central apnea. Compared to CRM data, sensitivity of STOP-BANG questionnaire in diagnosis of sleep-disordered breathing was 92% (specificity 33%). Sensitivity of ESS was 44%. Sensitivity of computer-based pulse oximetry for detecting sleep-disordered breathing was 100%. In 3 cases, computer-based pulse oximetry substantially underestimated severity of sleep-disordered breathing compared to CRM. Overestimation was noted in other 3 cases. In all 7 patients in whom respiratory disturbance indices were obtained from ICDs, mean overnight values indicated sleep-disordered breathing that was confirmed by CRM. In 3 patients with mild SDB confirmed by CRM, apnea detection algorithm in ICD markedly overestimated respiratory event index. Conclusion. In patients with ICDs, SDB is diagnosed in 95.5% of cases with moderate-to-severe forms identified in 61% patients. Compared to CRM data, STOP-BANG and ESS scales in patients with ICDs are not reliable in identifying SDB and cannot be recommended as the main screening methods. High sensitivity of computer-based pulse oximetry in detecting SDB, when compared with CRM data, supports recommendation of this method for screening in patients with ICDs.

Authors

Institutions

Publication Details

Journal
Russian Cardiology Bulletin
Published
2025-12-25
DOI
https://doi.org/10.17116/cardiobulletin20252004156
Primary Topic
Obstructive Sleep Apnea Research
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Screening capabilities and prevalence of sleep-disordered breathing in patients with implantable cardioverter-defibrillators

Н. Б. Шлевков, E M Elfimova, I.I. Ksanaev, A Yu Litvin et al.
Russian Cardiology Bulletin
Obstructive Sleep Apnea Research
article

Screening capabilities and prevalence of sleep-disordered breathing in patients with implantable cardioverter-defibrillators

Н. Б. Шлевков, E M Elfimova, I.I. Ksanaev, A Yu Litvin, O.O. Mikhailova, V.G. Kiktev, A.V. Pevzner
article en

Abstract

Objective. To assess diagnostic efficacy of screening methods for sleep-disordered breathing — questionnaires (assessing the risk of obstructive sleep apnea and daytime sleepiness), overnight pulse oximetry and apnea detection algorithm embedded in implantable cardioverter-defibrillators (ICDs) in patients with ICDs. Material and methods. The study included 67 patients aged 34-83 years (median 65). There were 54 (81%) men, median body mass index was 29.4 kg/m2. All patients underwent CRM to detect sleep-disordered breathing. Screening performance of various diagnostic methods for sleep apnea detection was assessed in 39 patients using STOP-Bang questionnaire and Epworth Sleepiness Scale (ESS), in 20 patients — using computer-based pulse oximetry (PO), in 7 patients — via interpretation of respiratory disturbance index obtained from ICDs of specific models. Results of these screening methods were compared with CRM data. Results. According to CRM data, sleep-disordered breathing was detected in 64 (95.5%) patients (mild in 23 (34.5%), moderate in 22 (33%), and severe in 19 (28%) cases). In all cases, sleep-disordered breathing was obstructive, but 19 (28%) patients had combination with significant (more than 5 events/hour) episodes of central apnea. Compared to CRM data, sensitivity of STOP-BANG questionnaire in diagnosis of sleep-disordered breathing was 92% (specificity 33%). Sensitivity of ESS was 44%. Sensitivity of computer-based pulse oximetry for detecting sleep-disordered breathing was 100%. In 3 cases, computer-based pulse oximetry substantially underestimated severity of sleep-disordered breathing compared to CRM. Overestimation was noted in other 3 cases. In all 7 patients in whom respiratory disturbance indices were obtained from ICDs, mean overnight values indicated sleep-disordered breathing that was confirmed by CRM. In 3 patients with mild SDB confirmed by CRM, apnea detection algorithm in ICD markedly overestimated respiratory event index. Conclusion. In patients with ICDs, SDB is diagnosed in 95.5% of cases with moderate-to-severe forms identified in 61% patients. Compared to CRM data, STOP-BANG and ESS scales in patients with ICDs are not reliable in identifying SDB and cannot be recommended as the main screening methods. High sensitivity of computer-based pulse oximetry in detecting SDB, when compared with CRM data, supports recommendation of this method for screening in patients with ICDs.

Russian Cardiology BulletinVol. 20(4)
National Medical Research Center of Cardiology (RU)
Good health and well-being
Openalex Percentile: Top 33%
Obstructive Sleep Apnea Research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.