Effect of vibrational forces from acceledent on dentoalveolar surface temperature during clear aligner therapy: an infrared thermography study

Mechanical vibration devices such as AcceleDent have been marketed as non-invasive adjuncts capable of accelerating orthodontic tooth movement. The proposed mechanism involves the stimulation of cellular activity and bone remodeling. The aim of this study was to determine whether the use of AcceleDent, in combination with clear aligner therapy, produces a measurable increase in the surface temperature of intraoral tissues. Twenty-two adult patients (mean age: 39 ± 15 years) undergoing orthodontic treatment with clear aligners (Invisalign) participated in this crossover study. Infrared thermography (A35, FLIR, Wilsonville, USA) was used to record temperature measurements of gingival and dental surfaces. Each participant underwent two experimental conditions on separate days: (1) wearing the AcceleDent device while active (vibrating) and (2) wearing the same device inactive (non-vibrating). After a 20-minute thermal acclimatization period, the device was applied for 20 min before imaging. No statistically significant differences were observed in surface temperature between the vibrating and non-vibrating conditions (Gingiva: p = 0.69; Tooth: p = 0.76). Likewise, no significant differences were detected at baseline between the two sessions (Gingiva: p = 0.35; Tooth: p = 0.57). The use of AcceleDent vibration during clear aligner therapy does not result in an increase in the surface temperature of gingival or dental tissues, as measured by infrared thermography. This study provides evidence that AcceleDent vibration does not produce thermal changes in intraoral tissues, suggesting it is thermally safe for use alongside clear aligners. Infrared thermography may serve as a non-invasive tool to monitor physiological effects of orthodontic devices.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2025-12-24
DOI
https://doi.org/10.1038/s41598-025-28065-8
Primary Topic
Infrared Thermography in Medicine
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Effect of vibrational forces from acceledent on dentoalveolar surface temperature during clear aligner therapy: an infrared thermography study

Jaime Gil López-Areal, B. Tarazona-Álvarez, José Luís Gandía-Franco, Andrea Vítores-Calero et al.
Scientific Reports
Infrared Thermography in Medicine
article

Effect of vibrational forces from acceledent on dentoalveolar surface temperature during clear aligner therapy: an infrared thermography study

Jaime Gil López-Areal, B. Tarazona-Álvarez, José Luís Gandía-Franco, Andrea Vítores-Calero, Jose Ignacio Priego-Quesada, Rosa Cibrián-Ortiz de Anda, Carlos Bellot-Arcís, Natalia Zamora-Martínez, Vanessa Paredes-Gallardo, Verónica García-Sanz
article en

Abstract

Mechanical vibration devices such as AcceleDent have been marketed as non-invasive adjuncts capable of accelerating orthodontic tooth movement. The proposed mechanism involves the stimulation of cellular activity and bone remodeling. The aim of this study was to determine whether the use of AcceleDent, in combination with clear aligner therapy, produces a measurable increase in the surface temperature of intraoral tissues. Twenty-two adult patients (mean age: 39 ± 15 years) undergoing orthodontic treatment with clear aligners (Invisalign) participated in this crossover study. Infrared thermography (A35, FLIR, Wilsonville, USA) was used to record temperature measurements of gingival and dental surfaces. Each participant underwent two experimental conditions on separate days: (1) wearing the AcceleDent device while active (vibrating) and (2) wearing the same device inactive (non-vibrating). After a 20-minute thermal acclimatization period, the device was applied for 20 min before imaging. No statistically significant differences were observed in surface temperature between the vibrating and non-vibrating conditions (Gingiva: p = 0.69; Tooth: p = 0.76). Likewise, no significant differences were detected at baseline between the two sessions (Gingiva: p = 0.35; Tooth: p = 0.57). The use of AcceleDent vibration during clear aligner therapy does not result in an increase in the surface temperature of gingival or dental tissues, as measured by infrared thermography. This study provides evidence that AcceleDent vibration does not produce thermal changes in intraoral tissues, suggesting it is thermally safe for use alongside clear aligners. Infrared thermography may serve as a non-invasive tool to monitor physiological effects of orthodontic devices.

Scientific ReportsVol. 15(1)
Universitat de València (ES)
Openalex Percentile: Top 31%
Infrared Thermography in Medicine
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.