Comparison of scandium-based post-plasma chemical ionization approaches for elemental fluorine detection by nitrogen microwave-sustained inductively-coupled atmospheric-pressure plasma

Fluorine detection has been a challenge in elemental mass spectrometry because of inefficient ion generation. Polyatomic analytical ion formation has emerged as a promising strategy to address this limitation. Here, we compare several post-plasma ion-neutral reactions that leverage affinity between scandium and F atoms to form analytical ions. A nitrogen microwave-sustained inductively-coupled atmospheric-pressure plasma (MICAP) is utilized to decompose fluorochemicals. The plasma products are sampled into an interface designed to minimize their adsorption to the surfaces, and are ionized in two main ways: 1) by scandium-containing ions generated using nano-electrospray ionization (nESI); 2) by addition of scandium to the plasma coupled with a post-MICAP corona discharge. Both avenues are also evaluated in positive- and negative-ion modes, producing ions such as ScFNO<sub>3</sub>(H<sub>2</sub>O)<sub>n</sub> <sup>+</sup> and ScF(NO<sub>3</sub>)<sub>3</sub> <sup>-</sup>. In positive-ion mode, both approaches show similar fundamental characteristics, suggesting HF ionization as the main signal generation pathway. However, in negative-ion mode, nESI shows significantly lower ionization efficiency compared to that of the post-plasma corona discharge approach. The better ionization efficiency by negative-mode corona discharge ionization is attributed to production of more effective reagent ions for HF ionization, and possibility of ionizing Sc- and F-containing neutral plasma products. Importantly, the isobaric interferences are significantly decreased in negative-ion mode, improving analytical performance. Further, negative-mode ionization offers better tolerance to sodium matrix compared to that of positive-mode ionization, attributed to decreased clustering tendencies of negative ions with plasma-produced NaNO<sub>3</sub> upon sodium introduction. These studies provide critical insights to further improve post-plasma chemical ionization for F detection.

Authors

Institutions

Publication Details

Journal
Spectrochimica Acta Part B Atomic Spectroscopy
Published
2025-12-24
DOI
https://doi.org/10.1016/j.sab.2025.107311
Primary Topic
Mass Spectrometry Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparison of scandium-based post-plasma chemical ionization approaches for elemental fluorine detection by nitrogen microwave-sustained inductively-coupled atmospheric-pressure plasma

Kaveh Jorabchi, Jordan L. Tanen
Spectrochimica Acta Part B Atomic Spectroscopy
Mass Spectrometry Techniques and Applications
article

Comparison of scandium-based post-plasma chemical ionization approaches for elemental fluorine detection by nitrogen microwave-sustained inductively-coupled atmospheric-pressure plasma

Kaveh Jorabchi, Jordan L. Tanen
article en

Abstract

Fluorine detection has been a challenge in elemental mass spectrometry because of inefficient ion generation. Polyatomic analytical ion formation has emerged as a promising strategy to address this limitation. Here, we compare several post-plasma ion-neutral reactions that leverage affinity between scandium and F atoms to form analytical ions. A nitrogen microwave-sustained inductively-coupled atmospheric-pressure plasma (MICAP) is utilized to decompose fluorochemicals. The plasma products are sampled into an interface designed to minimize their adsorption to the surfaces, and are ionized in two main ways: 1) by scandium-containing ions generated using nano-electrospray ionization (nESI); 2) by addition of scandium to the plasma coupled with a post-MICAP corona discharge. Both avenues are also evaluated in positive- and negative-ion modes, producing ions such as ScFNO<sub>3</sub>(H<sub>2</sub>O)<sub>n</sub> <sup>+</sup> and ScF(NO<sub>3</sub>)<sub>3</sub> <sup>-</sup>. In positive-ion mode, both approaches show similar fundamental characteristics, suggesting HF ionization as the main signal generation pathway. However, in negative-ion mode, nESI shows significantly lower ionization efficiency compared to that of the post-plasma corona discharge approach. The better ionization efficiency by negative-mode corona discharge ionization is attributed to production of more effective reagent ions for HF ionization, and possibility of ionizing Sc- and F-containing neutral plasma products. Importantly, the isobaric interferences are significantly decreased in negative-ion mode, improving analytical performance. Further, negative-mode ionization offers better tolerance to sodium matrix compared to that of positive-mode ionization, attributed to decreased clustering tendencies of negative ions with plasma-produced NaNO<sub>3</sub> upon sodium introduction. These studies provide critical insights to further improve post-plasma chemical ionization for F detection.

Spectrochimica Acta Part B Atomic SpectroscopyVol. 236
Georgetown University (US)
Openalex Percentile: Top 40%
Mass Spectrometry Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.