Size detection limits in single particle inductively coupled plasma mass spectrometry: Reconsidering the selection of dwell times

Performance of single particle inductively coupled plasma mass spectrometry (SP-ICP-MS) methods is constrained by the characteristics of the instrumentation. The commercial availability of fast data acquisition quadrupole instruments, which allow the use of dwell times in the microsecond range rather than being restricted to milliseconds, was expected to improve the performance of the technique. When data acquisition frequency is increased, individual particles are recorded as transient signals consisting of multiple-reading events instead of one-reading pulses. Since particle events must be detected above the baseline noise, the highest intensity reading of a transient signal becomes a relevant parameter whose value decreases with decreasing dwell time, as the total intensity of a particle event is independent of the number of readings per event and thus of the selected dwell time. In this work, the effect of dwell time on the attainable size detection limits has been reconsidered, achieving minimum size detection limits when dwell times in the range of 200 – 500 μs were used, regardless of the baseline level. At such dwell times, particle events from small nanoparticles (e.g., 2 – 3 times the size detection limit) were recorded within 1 – 2 readings despite the duration of particle events, which was modified working in both standard and gas reaction modes. Under these conditions, free and proprietary software capable of handling multiple-reading events allowed to process successfully the SP-ICP-MS data independently of the dwell time applied. • The lowest size detection limits in SP-ICP-MS are achieved at 200–500 μs dwell times. • At such dwell times, particle events are recorded within 1–3 readings. • Particles events can be successfully processed by using peak-based software. • The magnitude of the baseline does not affect to this behaviour.

Authors

Institutions

Publication Details

Journal
Spectrochimica Acta Part B Atomic Spectroscopy
Published
2025-12-24
DOI
https://doi.org/10.1016/j.sab.2025.107307
Primary Topic
Mass Spectrometry Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Size detection limits in single particle inductively coupled plasma mass spectrometry: Reconsidering the selection of dwell times

Eduardo Bolea, Isabel Abad-Álvaro, Francisco Laborda
Spectrochimica Acta Part B Atomic Spectroscopy
Mass Spectrometry Techniques and Applications
article

Size detection limits in single particle inductively coupled plasma mass spectrometry: Reconsidering the selection of dwell times

Eduardo Bolea, Isabel Abad-Álvaro, Francisco Laborda
article en

Abstract

Performance of single particle inductively coupled plasma mass spectrometry (SP-ICP-MS) methods is constrained by the characteristics of the instrumentation. The commercial availability of fast data acquisition quadrupole instruments, which allow the use of dwell times in the microsecond range rather than being restricted to milliseconds, was expected to improve the performance of the technique. When data acquisition frequency is increased, individual particles are recorded as transient signals consisting of multiple-reading events instead of one-reading pulses. Since particle events must be detected above the baseline noise, the highest intensity reading of a transient signal becomes a relevant parameter whose value decreases with decreasing dwell time, as the total intensity of a particle event is independent of the number of readings per event and thus of the selected dwell time. In this work, the effect of dwell time on the attainable size detection limits has been reconsidered, achieving minimum size detection limits when dwell times in the range of 200 – 500 μs were used, regardless of the baseline level. At such dwell times, particle events from small nanoparticles (e.g., 2 – 3 times the size detection limit) were recorded within 1 – 2 readings despite the duration of particle events, which was modified working in both standard and gas reaction modes. Under these conditions, free and proprietary software capable of handling multiple-reading events allowed to process successfully the SP-ICP-MS data independently of the dwell time applied. • The lowest size detection limits in SP-ICP-MS are achieved at 200–500 μs dwell times. • At such dwell times, particle events are recorded within 1–3 readings. • Particles events can be successfully processed by using peak-based software. • The magnitude of the baseline does not affect to this behaviour.

Spectrochimica Acta Part B Atomic SpectroscopyVol. 236
Universidad de Zaragoza (ES)
Quality Education
Openalex Percentile: Top 40%
Mass Spectrometry Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Size detection limits in single particle inductively coupled plasma mass spectrometry: Reconsidering the selection of dwell times — Eduardo Bolea, Isabel Abad-Álvaro, et al. · Spectrochimica Acta Part B Atomic Spectroscopy (2025) | TGRS Research Map | TGRS