Systematic Development and Validation of a Bradford-Based Protein Quantification Method for Novel Multi-Dose R21 Malaria Vaccine Formulated with 2-Phenoxy Ethanol (2-PE)

<b>Background:</b> The R21 malaria vaccine is a next-generation, WHO-prequalified vaccine that was introduced to reduce the burden of clinical malaria. In alignment with WHO recommendations, multi-dose vaccine presentations are preferred for large-scale immunization and inclusion in the Expanded Programme on Immunization (EPI). Accurate protein quantification is a critical quality control parameter for lot release, but it remains challenging when the antigen is present at low protein concentrations or formulated with complex matrices, including adjuvants, stabilizers, and preservatives. <b>Methods:</b> In this study, multiple protein estimation methods including Micro-BCA, BCA, and Bradford assays were evaluated to determine their suitability for quantifying the R21 antigen formulated with Matrix-M1 adjuvant and 2-PE preservative. The Bradford assay was selected as the most appropriate method, based on a comparative assessment of precision, accuracy, and linearity. Further optimization was undertaken to identify suitable buffer systems, and the method was validated in accordance with ICH Q2(R2) guidelines. <b>Results:</b> Validation results demonstrated that the assay is specific, accurate, precise, and repeatable, with a limit of quantitation (LOQ) of 2 µg/mL. The method demonstrated comparable performance to ELISA and was found to be sensitive enough to detect changes in antigen concentration resulting from unintended adsorption of R21 to vial surfaces. The assay offers a rapid, high-throughput, and cost-effective solution for protein quantitation in commercial manufacturing, lot release, and stability studies. The protein content of the drug product, quantified using the Bradford method, demonstrated robust in vivo immunogenicity in both release and stability studies. <b>Conclusions:</b> The robustness and reproducibility of the assay establish a new benchmark in quality control for virus-like particle (VLP)-based vaccines with complex formulations, thereby supporting the precision and reliability required for global malaria prevention efforts.

Authors

Institutions

Publication Details

Journal
Vaccines
Published
2025-12-24
DOI
https://doi.org/10.3390/vaccines14010025
Citations
2
Primary Topic
vaccines and immunoinformatics approaches
Type
article
Field-Weighted Citation Impact
1.61

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Systematic Development and Validation of a Bradford-Based Protein Quantification Method for Novel Multi-Dose R21 Malaria Vaccine Formulated with 2-Phenoxy Ethanol (2-PE)

Dnyanesh Ranade, Ajay T. Salunke, Rajender Jena, Sunil Gairola et al.
2 citations
Vaccines
vaccines and immunoinformatics approaches
1.61
article

Systematic Development and Validation of a Bradford-Based Protein Quantification Method for Novel Multi-Dose R21 Malaria Vaccine Formulated with 2-Phenoxy Ethanol (2-PE)

Dnyanesh Ranade, Ajay T. Salunke, Rajender Jena, Sunil Gairola, Prajwal Chaudhari, Aniket Mahamuni
article en
2 citations

Abstract

<b>Background:</b> The R21 malaria vaccine is a next-generation, WHO-prequalified vaccine that was introduced to reduce the burden of clinical malaria. In alignment with WHO recommendations, multi-dose vaccine presentations are preferred for large-scale immunization and inclusion in the Expanded Programme on Immunization (EPI). Accurate protein quantification is a critical quality control parameter for lot release, but it remains challenging when the antigen is present at low protein concentrations or formulated with complex matrices, including adjuvants, stabilizers, and preservatives. <b>Methods:</b> In this study, multiple protein estimation methods including Micro-BCA, BCA, and Bradford assays were evaluated to determine their suitability for quantifying the R21 antigen formulated with Matrix-M1 adjuvant and 2-PE preservative. The Bradford assay was selected as the most appropriate method, based on a comparative assessment of precision, accuracy, and linearity. Further optimization was undertaken to identify suitable buffer systems, and the method was validated in accordance with ICH Q2(R2) guidelines. <b>Results:</b> Validation results demonstrated that the assay is specific, accurate, precise, and repeatable, with a limit of quantitation (LOQ) of 2 µg/mL. The method demonstrated comparable performance to ELISA and was found to be sensitive enough to detect changes in antigen concentration resulting from unintended adsorption of R21 to vial surfaces. The assay offers a rapid, high-throughput, and cost-effective solution for protein quantitation in commercial manufacturing, lot release, and stability studies. The protein content of the drug product, quantified using the Bradford method, demonstrated robust in vivo immunogenicity in both release and stability studies. <b>Conclusions:</b> The robustness and reproducibility of the assay establish a new benchmark in quality control for virus-like particle (VLP)-based vaccines with complex formulations, thereby supporting the precision and reliability required for global malaria prevention efforts.

VaccinesVol. 14(1)
Serum Institute of India (India) (IN)
Novavax
Good health and well-being
Openalex Percentile: Top 14%
vaccines and immunoinformatics approaches
1.61
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.