In situ hydride generation for multielement analysis in complex matrices through ICP-MS with a high-temperature sample introduction system

Hydride generation (HG) remains a powerful sample introduction strategy for the determination of some elements by inductively coupled plasma mass spectrometry (ICP-MS). In this study, two HG-based configurations were systematically evaluated, including a conventional HG system with external gas-liquid separation and a structurally modified high-temperature torch-integrated sample introduction system (hTISIS). The hTISIS was equipped with an auxiliary inlet, i.e., drain tube, for the introduction of NaBH 4 and HCl, allowing gas-liquid reaction and phase separation to occur directly within the thermostated spray chamber. This latter setup, HG-hTISIS, was employed under three different modes: sample introduction through both the nebulizer and the drain tube (mode 1); sample introduction only through nebulization (mode 2); and sample introduction only through the drain tube (mode 3). In all three instances, hydride-forming reagents were continuously delivered by means of the drain tube. The operability of all configurations was assessed for five hydride-forming elements (As, Se, Sb, Sn, Bi), while quantitative performance was ultimately established using the optimized in situ HG-hTISIS configuration operated at 200 °C. Non-hydride-forming elements including Li, Al, Mn, Ti, Co, V, Cr, Ni, Cu, Zn, Rb, Sr, Ag, Ba, Cd, Tl, and Pb were also determined under this configuration, with an efficient aerosol transport enabled by the thermal chamber. A Box-Behnken design was employed to optimize the concentrations of NaOH, NaBH 4 , and HCl, ensuring robust hydride formation and minimal matrix effects across a wide range of conditions. Analytical performance was evaluated in terms of sensitivity, precision, and limits of detection (LODs). Signal enhancement factors ranging from 13 to 150 were achieved for hydride-forming elements, with sub-μg L −1 detection limits and a sample uptake of only 30 μL min −1 , in contrast to 0.5 mL min −1 in the conventional HG system. Moreover, foam formation was significantly mitigated in biological matrices, allowing direct analysis after sample dilution. Accuracy was confirmed through the analysis of certified reference materials of environmental and clinical origin, showing good agreement with the certified values. These results highlighted the analytical potential of the hydride generation HG-hTISIS configuration as a sensitive, low-consumption, and multielement-compatible alternative for the analysis of complex matrices by ICP-MS. • A modified high-temperature torch integrated system has been adapted for hydride generation ICP-MS analysis. • Hydride and non-hydride generation elements have been determined in clinical and environmental CRMs through the HG-hTISIS. • The HG-hTISIS provides low enough limits of detection and accurate results for CRMs.

Authors

Publication Details

Journal
Spectrochimica Acta Part B Atomic Spectroscopy
Published
2025-12-24
DOI
https://doi.org/10.1016/j.sab.2025.107351
Primary Topic
Analytical chemistry methods development
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

In situ hydride generation for multielement analysis in complex matrices through ICP-MS with a high-temperature sample introduction system

José Luis Todolí-Torró, Rebeca Pérez-Ramírez, Raquel Sánchez-Romero
Spectrochimica Acta Part B Atomic Spectroscopy
Analytical chemistry methods development
article

In situ hydride generation for multielement analysis in complex matrices through ICP-MS with a high-temperature sample introduction system

José Luis Todolí-Torró, Rebeca Pérez-Ramírez, Raquel Sánchez-Romero
article en

Abstract

Hydride generation (HG) remains a powerful sample introduction strategy for the determination of some elements by inductively coupled plasma mass spectrometry (ICP-MS). In this study, two HG-based configurations were systematically evaluated, including a conventional HG system with external gas-liquid separation and a structurally modified high-temperature torch-integrated sample introduction system (hTISIS). The hTISIS was equipped with an auxiliary inlet, i.e., drain tube, for the introduction of NaBH 4 and HCl, allowing gas-liquid reaction and phase separation to occur directly within the thermostated spray chamber. This latter setup, HG-hTISIS, was employed under three different modes: sample introduction through both the nebulizer and the drain tube (mode 1); sample introduction only through nebulization (mode 2); and sample introduction only through the drain tube (mode 3). In all three instances, hydride-forming reagents were continuously delivered by means of the drain tube. The operability of all configurations was assessed for five hydride-forming elements (As, Se, Sb, Sn, Bi), while quantitative performance was ultimately established using the optimized in situ HG-hTISIS configuration operated at 200 °C. Non-hydride-forming elements including Li, Al, Mn, Ti, Co, V, Cr, Ni, Cu, Zn, Rb, Sr, Ag, Ba, Cd, Tl, and Pb were also determined under this configuration, with an efficient aerosol transport enabled by the thermal chamber. A Box-Behnken design was employed to optimize the concentrations of NaOH, NaBH 4 , and HCl, ensuring robust hydride formation and minimal matrix effects across a wide range of conditions. Analytical performance was evaluated in terms of sensitivity, precision, and limits of detection (LODs). Signal enhancement factors ranging from 13 to 150 were achieved for hydride-forming elements, with sub-μg L −1 detection limits and a sample uptake of only 30 μL min −1 , in contrast to 0.5 mL min −1 in the conventional HG system. Moreover, foam formation was significantly mitigated in biological matrices, allowing direct analysis after sample dilution. Accuracy was confirmed through the analysis of certified reference materials of environmental and clinical origin, showing good agreement with the certified values. These results highlighted the analytical potential of the hydride generation HG-hTISIS configuration as a sensitive, low-consumption, and multielement-compatible alternative for the analysis of complex matrices by ICP-MS. • A modified high-temperature torch integrated system has been adapted for hydride generation ICP-MS analysis. • Hydride and non-hydride generation elements have been determined in clinical and environmental CRMs through the HG-hTISIS. • The HG-hTISIS provides low enough limits of detection and accurate results for CRMs.

Spectrochimica Acta Part B Atomic SpectroscopyVol. 236
Openalex Percentile: Top 42%
Analytical chemistry methods development
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.