Validation of the RBP‐9000c Oscillometric Blood Pressure Monitor in the General Population According to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/ International Organization for Standardization Un

The aim of this study was to evaluate the accuracy of the single upper-arm cuff oscillometric blood pressure (BP) monitor RBP-9000 c developed for office and home blood pressure measurement in the general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2:2018). Subjects were recruited to fulfill the age, gender, BP, and cuff distribution criteria of the AAMI/ESH/ISO Universal Standard in the general population using the same-arm sequential BP measurement method. The test device incorporates a single built-in cuff suitable for 17-42 cm arm circumference. For validation criterion 1, the mean ± SD of the differences between the test device and reference BP readings was 2.4 ± 6.7/3.3 ± 6.3 mmHg (systolic/ diastolic). For criterion 2, the SD of the mean BP differences between the test device and reference BP per subject was 5.28/5.32 mmHg (systolic/diastolic). The RBP-9000c oscillometric device for office and home BP measurement fulfilled all the requirements of the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018) in the general population and can be recommended for clinical and self-use at home. Trial Registration: ChiCTR2300075747.

Authors

Institutions

Publication Details

Journal
Journal of Clinical Hypertension
Published
2025-12-01
DOI
https://doi.org/10.1111/jch.70194
Primary Topic
Blood Pressure and Hypertension Studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Validation of the RBP‐9000c Oscillometric Blood Pressure Monitor in the General Population According to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/ International Organization for Standardization Un

Shijie Yang, Zhanyang Zhou, Miao He, Yuqing Zhang
Journal of Clinical Hypertension
Blood Pressure and Hypertension Studies
article

Validation of the RBP‐9000c Oscillometric Blood Pressure Monitor in the General Population According to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/ International Organization for Standardization Un

Shijie Yang, Zhanyang Zhou, Miao He, Yuqing Zhang
article en

Abstract

The aim of this study was to evaluate the accuracy of the single upper-arm cuff oscillometric blood pressure (BP) monitor RBP-9000 c developed for office and home blood pressure measurement in the general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2:2018). Subjects were recruited to fulfill the age, gender, BP, and cuff distribution criteria of the AAMI/ESH/ISO Universal Standard in the general population using the same-arm sequential BP measurement method. The test device incorporates a single built-in cuff suitable for 17-42 cm arm circumference. For validation criterion 1, the mean ± SD of the differences between the test device and reference BP readings was 2.4 ± 6.7/3.3 ± 6.3 mmHg (systolic/ diastolic). For criterion 2, the SD of the mean BP differences between the test device and reference BP per subject was 5.28/5.32 mmHg (systolic/diastolic). The RBP-9000c oscillometric device for office and home BP measurement fulfilled all the requirements of the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018) in the general population and can be recommended for clinical and self-use at home. Trial Registration: ChiCTR2300075747.

Journal of Clinical HypertensionVol. 27(12)
Chinese Academy of Medical Sciences & Peking Union Medical College (CN), Beijing Anzhen Hospital (CN), Fu Wai Hospital (CN)
Openalex Percentile: Top 31%
Blood Pressure and Hypertension Studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.