An in vitro assessment of aspiration technologies: Location of conduit for 6F aspiration catheters and design of catheter tip influence clot retrieval success

Background8F catheters serve as conduits to provide stable access and support for delivery of 6F aspiration catheters (ACs) in Acute Ischaemic Stroke treatment. Superbore catheters (SBCs) (internal diameter (ID) > 0.088″) are the latest innovation, designed for placement in the middle cerebral artery (MCA). SBCs may improve recanalization via enhanced control of blood pressure and flow, in comparison to balloon guide catheter (BGC) and regular guide catheter (GC) systems.AimTo objectively assess conduit catheter technologies in terms of deliverability, local impact on blood pressure and flow, and clot retrieval success in vitro.MethodsTwo in vitro models were used. Two SBCs, (CEREGLIDE92 and TracStar/Zoom88) and one BGC (EMBOGUARD) were used with 0.071″ ACs, and one GC (NeuronMAX) was evaluated with two 6F ACs (RED68 and SOFIA). M1-MCA occlusions were created with cohesive and friable thrombus analogues. Device tracking and clot retrieval were evaluated by two neurointerventionalists, under physiological haemodynamic conditions.ResultsSBC positioning in the MCA resulted in significant reductions of local blood pressure and flow in comparison to GC systems positioned in the Internal Carotid Artery (<i>p</i> = .000). However, when the SBC could not reach the MCA, they displayed similar haemodynamic control as the GC systems. SBC systems achieved high retrieval success for friable thrombi, likely due to flow reversal in the MCA during aspiration (<i>p</i> = .000). The TracStar/Zoom SBC system had lower rates of cohesive retrieval success due to difficulties in catheter tracking and poor clot alignment with the catheter bevelled tip.ConclusionClot retrieval success is influenced by the location of the conduit for the 6F AC and the design of the AC tip.

Authors

Institutions

Publication Details

Journal
Interventional Neuroradiology
Published
2025-12-23
DOI
https://doi.org/10.1177/15910199251405913
Primary Topic
Cerebrospinal fluid and hydrocephalus
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

An in vitro assessment of aspiration technologies: Location of conduit for 6F aspiration catheters and design of catheter tip influence clot retrieval success

Tommy Andersson, Bruce P. Murphy, Aoife Glynn, Caitríona Lally et al.
Interventional Neuroradiology
Cerebrospinal fluid and hydrocephalus
article

An in vitro assessment of aspiration technologies: Location of conduit for 6F aspiration catheters and design of catheter tip influence clot retrieval success

Tommy Andersson, Bruce P. Murphy, Aoife Glynn, Caitríona Lally, Ray McCarthy, AnnaLisa Smullin, Sarah Power
article en

Abstract

Background8F catheters serve as conduits to provide stable access and support for delivery of 6F aspiration catheters (ACs) in Acute Ischaemic Stroke treatment. Superbore catheters (SBCs) (internal diameter (ID) > 0.088″) are the latest innovation, designed for placement in the middle cerebral artery (MCA). SBCs may improve recanalization via enhanced control of blood pressure and flow, in comparison to balloon guide catheter (BGC) and regular guide catheter (GC) systems.AimTo objectively assess conduit catheter technologies in terms of deliverability, local impact on blood pressure and flow, and clot retrieval success in vitro.MethodsTwo in vitro models were used. Two SBCs, (CEREGLIDE92 and TracStar/Zoom88) and one BGC (EMBOGUARD) were used with 0.071″ ACs, and one GC (NeuronMAX) was evaluated with two 6F ACs (RED68 and SOFIA). M1-MCA occlusions were created with cohesive and friable thrombus analogues. Device tracking and clot retrieval were evaluated by two neurointerventionalists, under physiological haemodynamic conditions.ResultsSBC positioning in the MCA resulted in significant reductions of local blood pressure and flow in comparison to GC systems positioned in the Internal Carotid Artery (<i>p</i> = .000). However, when the SBC could not reach the MCA, they displayed similar haemodynamic control as the GC systems. SBC systems achieved high retrieval success for friable thrombi, likely due to flow reversal in the MCA during aspiration (<i>p</i> = .000). The TracStar/Zoom SBC system had lower rates of cohesive retrieval success due to difficulties in catheter tracking and poor clot alignment with the catheter bevelled tip.ConclusionClot retrieval success is influenced by the location of the conduit for the 6F AC and the design of the AC tip.

Interventional Neuroradiology
Karolinska University Hospital (SE), Trinity College Dublin (IE), AZ Groeninge (BE), Botsford Hospital (US), Johnson & Johnson (Ireland) (IE), Beaumont Hospital, Dearborn (US)
Industry, innovation and infrastructure
Openalex Percentile: Top 38%
Cerebrospinal fluid and hydrocephalus
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.