Accuracy of a continuous glucose monitoring system in people with diabetes on peritoneal dialysis

BackgroundLack of data on the accuracy of continuous glucose monitoring systems currently limits their usage in people with diabetes on peritoneal dialysis (PD).AimWe aimed to assess the analytical and clinical accuracy of the FreeStyle Libre (FSL) continuous glucose monitoring system in people with diabetes on PD.Methods12 participants using Icodextrin in their PD regime were recruited to a single-centre observational study. They wore a blinded research model of the FSL for 10 days. Results from the FSL were compared with venous glucose measured on a Yellow Springs Instrument (YSI) and self-monitored capillary blood glucose (SMBG) recorded five times per day over the 10-day study.ResultsThe mean absolute relative difference from 84 FSL-YSI matched pairs was 9.8% (95% CI 8.6-11.1) and from 416 FSL-SMBG matched pairs, it was 17.3% (95% CI 16.24-18.43). The systematic error for the FSL as determined by Bland-Altman analysis was -0.6 ± 1.0 mmol/l compared with YSI and -1.4 ± 1.9 mmol/l compared with SMBG. With regard clinical accuracy, compared with YSI and SMBG, respectively, 100% and 99.9% of sensor values were in clinically acceptable zones A and B of Parkes consensus error grid.ConclusionWe demonstrated satisfactory performance of the FSL monitoring system by both analytical and clinical metrics in this cohort of PD patients using treatment prescriptions including Icodextrin-based fluids. Larger studies are now needed to provide clinicians with appropriate reassurance if this technology is to be used with confidence in people on PD.

Authors

Institutions

Publication Details

Journal
Peritoneal Dialysis International
Published
2025-12-23
DOI
https://doi.org/10.1177/08968608251406514
Primary Topic
Diabetes Management and Research
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Accuracy of a continuous glucose monitoring system in people with diabetes on peritoneal dialysis

Mark Gilchrist, Jennifer Williams, Angela C. Shore
Peritoneal Dialysis International
Diabetes Management and Research
article

Accuracy of a continuous glucose monitoring system in people with diabetes on peritoneal dialysis

Mark Gilchrist, Jennifer Williams, Angela C. Shore
article en

Abstract

BackgroundLack of data on the accuracy of continuous glucose monitoring systems currently limits their usage in people with diabetes on peritoneal dialysis (PD).AimWe aimed to assess the analytical and clinical accuracy of the FreeStyle Libre (FSL) continuous glucose monitoring system in people with diabetes on PD.Methods12 participants using Icodextrin in their PD regime were recruited to a single-centre observational study. They wore a blinded research model of the FSL for 10 days. Results from the FSL were compared with venous glucose measured on a Yellow Springs Instrument (YSI) and self-monitored capillary blood glucose (SMBG) recorded five times per day over the 10-day study.ResultsThe mean absolute relative difference from 84 FSL-YSI matched pairs was 9.8% (95% CI 8.6-11.1) and from 416 FSL-SMBG matched pairs, it was 17.3% (95% CI 16.24-18.43). The systematic error for the FSL as determined by Bland-Altman analysis was -0.6 ± 1.0 mmol/l compared with YSI and -1.4 ± 1.9 mmol/l compared with SMBG. With regard clinical accuracy, compared with YSI and SMBG, respectively, 100% and 99.9% of sensor values were in clinically acceptable zones A and B of Parkes consensus error grid.ConclusionWe demonstrated satisfactory performance of the FSL monitoring system by both analytical and clinical metrics in this cohort of PD patients using treatment prescriptions including Icodextrin-based fluids. Larger studies are now needed to provide clinicians with appropriate reassurance if this technology is to be used with confidence in people on PD.

Peritoneal Dialysis International
Royal Devon & Exeter NHS Foundation Trust (GB), Royal Devon and Exeter Hospital (GB), NIHR Exeter Clinical Research Facility (GB), Exeter Hospital (GB)
Good health and well-being
Openalex Percentile: Top 31%
Diabetes Management and Research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.