Development of a multi-parametric MRI platform to evaluate steady-state water content and optical properties in multiple ex vivo bovine lens

Here we describe the development of magnetic resonance imaging (MRI) protocols which when run on a standard clinical 3T MRI system can non-invasively monitor changes in the water content and refractive properties of bovine lenses maintained under organ culture. Fresh bovine lenses were incubated in artificial aqueous humour and placed in a custom holder within a hand/wrist coil. Multi-parametric mapping sequences (T1, T2∗, proton density [PD], and magnetisation transfer [MT] ratio) were optimised to quantify the total, free, and bound water compartments in distinct lens regions (outer cortex, inner cortex, and nucleus). Repeated scanning on different days demonstrated robust inter-sample reproducibility. MRI-derived T2∗ and MT ratio values were calibrated against refractive index (n) profiles calculated using laser ray tracing and subsequently applied to ZEMAX optical modelling to calculate lens power. The imaging protocol achieved high spatial resolution and reproducibility, which allowed the detection of regional variations in water composition within a practical scan time. Hypotonic lens swelling induced significant increases in total and free water across all lens regions but had no effect on bound water or the refractive index gradient. Lens swelling increased lens power (ΔP = 2.98D, p = 0.04) primarily due to a change in the surface geometry of the lens. This ability to quantify lens water states and refractive changes across multiple lens samples will facilitate the screening of novel pharmacological reagents that modulate lens water transport, allowing for testing of their efficacy as potential anti-cataract therapies.

Authors

Institutions

Publication Details

Journal
Experimental Eye Research
Published
2025-12-23
DOI
https://doi.org/10.1016/j.exer.2025.110822
Primary Topic
Connexins and lens biology
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Development of a multi-parametric MRI platform to evaluate steady-state water content and optical properties in multiple ex vivo bovine lens

Paul J. Donaldson, Thomas W. White, Catherine Morgan, Yadi Chen et al.
Experimental Eye Research
Connexins and lens biology
article

Development of a multi-parametric MRI platform to evaluate steady-state water content and optical properties in multiple ex vivo bovine lens

Paul J. Donaldson, Thomas W. White, Catherine Morgan, Yadi Chen, Beau Pontré, Jin Jin, Courtney A. Thorne, Chen Qiu, Xingzheng Pan
article en

Abstract

Here we describe the development of magnetic resonance imaging (MRI) protocols which when run on a standard clinical 3T MRI system can non-invasively monitor changes in the water content and refractive properties of bovine lenses maintained under organ culture. Fresh bovine lenses were incubated in artificial aqueous humour and placed in a custom holder within a hand/wrist coil. Multi-parametric mapping sequences (T1, T2∗, proton density [PD], and magnetisation transfer [MT] ratio) were optimised to quantify the total, free, and bound water compartments in distinct lens regions (outer cortex, inner cortex, and nucleus). Repeated scanning on different days demonstrated robust inter-sample reproducibility. MRI-derived T2∗ and MT ratio values were calibrated against refractive index (n) profiles calculated using laser ray tracing and subsequently applied to ZEMAX optical modelling to calculate lens power. The imaging protocol achieved high spatial resolution and reproducibility, which allowed the detection of regional variations in water composition within a practical scan time. Hypotonic lens swelling induced significant increases in total and free water across all lens regions but had no effect on bound water or the refractive index gradient. Lens swelling increased lens power (ΔP = 2.98D, p = 0.04) primarily due to a change in the surface geometry of the lens. This ability to quantify lens water states and refractive changes across multiple lens samples will facilitate the screening of novel pharmacological reagents that modulate lens water transport, allowing for testing of their efficacy as potential anti-cataract therapies.

Experimental Eye ResearchVol. 264
University of Auckland (NZ), Auckland UniServices (New Zealand) (NZ), Siemens Healthcare (United States) (US), Stony Brook University (US)
Clean water and sanitation
Openalex Percentile: Top 37%
Connexins and lens biology
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies