Preliminary investigation into the use of “micro-EVLA” endovenous surgery for bulging aesthetic veins using a 1940 nm diode laser and 272-micron bare fibre in a porcine liver model

AimsProminent or bulging veins on the temple or around the eyes are usually treated with extra-corporeal laser. However, when too big for laser to be effective, other techniques need to be used. The aim of this study was to evaluate the potential of using an endovenous approach to treat these small veins - a technique that we call "micro-EVLA" (micro-endovenous laser ablation).MethodsWe decided to use a 272 µm bare laser fibre, which we could pass through a 21G "green" hypodermic needle. In addition, we used a 1940 nm diode laser, as we have previously shown that the increased affinity of water to this wavelength results in a reduced thermal spread in biological tissues at low powers. We performed two sequential studies. The first measured the thermal spread in the porcine liver model using 1W, 2W and 3W, with a pull-back rate of either 5 s/cm or 10 s/cm. Two blinded observers measured the total width of the ablation tract and the spread of tissue ablation from the edge of the fibre. The second assessed the temperature of the liver surface using an infrared camera, above the laser fibre during treatment, with and without cold air cooling.ResultsIncreased power resulted in significantly increased thermal spread, as did the slower pull-back compared to the faster. We found that 2W at 5 s/cm (LEED 10 J/cm) appeared to give the optimal ablation with 0.20 mm (SD: 0.073) thermal spread from the edge of the fibre. When these settings were used with the fibre positioned 1.0-2.0 mm under the surface of the liver, cold air stopped any significant heating of the overlying liver surface.ConclusionOur micro-EVLA technique produced a controlled ablation tract, which appears to avoid the risk of skin burns. Clinical studies are required to confirm this conclusion.

Authors

Institutions

Publication Details

Journal
Phlebology The Journal of Venous Disease
Published
2025-12-23
DOI
https://doi.org/10.1177/02683555251411437
Primary Topic
Diagnosis and Treatment of Venous Diseases
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Preliminary investigation into the use of “micro-EVLA” endovenous surgery for bulging aesthetic veins using a 1940 nm diode laser and 272-micron bare fibre in a porcine liver model

Simon D Muschamp, Mark Steven Whiteley, Gabriel Mark Placzek, Sienna Esme Davey et al.
Phlebology The Journal of Venous Disease
Diagnosis and Treatment of Venous Diseases
article

Preliminary investigation into the use of “micro-EVLA” endovenous surgery for bulging aesthetic veins using a 1940 nm diode laser and 272-micron bare fibre in a porcine liver model

Simon D Muschamp, Mark Steven Whiteley, Gabriel Mark Placzek, Sienna Esme Davey, Oscar F Leveque
article en

Abstract

AimsProminent or bulging veins on the temple or around the eyes are usually treated with extra-corporeal laser. However, when too big for laser to be effective, other techniques need to be used. The aim of this study was to evaluate the potential of using an endovenous approach to treat these small veins - a technique that we call "micro-EVLA" (micro-endovenous laser ablation).MethodsWe decided to use a 272 µm bare laser fibre, which we could pass through a 21G "green" hypodermic needle. In addition, we used a 1940 nm diode laser, as we have previously shown that the increased affinity of water to this wavelength results in a reduced thermal spread in biological tissues at low powers. We performed two sequential studies. The first measured the thermal spread in the porcine liver model using 1W, 2W and 3W, with a pull-back rate of either 5 s/cm or 10 s/cm. Two blinded observers measured the total width of the ablation tract and the spread of tissue ablation from the edge of the fibre. The second assessed the temperature of the liver surface using an infrared camera, above the laser fibre during treatment, with and without cold air cooling.ResultsIncreased power resulted in significantly increased thermal spread, as did the slower pull-back compared to the faster. We found that 2W at 5 s/cm (LEED 10 J/cm) appeared to give the optimal ablation with 0.20 mm (SD: 0.073) thermal spread from the edge of the fibre. When these settings were used with the fibre positioned 1.0-2.0 mm under the surface of the liver, cold air stopped any significant heating of the overlying liver surface.ConclusionOur micro-EVLA technique produced a controlled ablation tract, which appears to avoid the risk of skin burns. Clinical studies are required to confirm this conclusion.

Phlebology The Journal of Venous Disease
Whiteley Clinic (GB)
Openalex Percentile: Top 27%
Diagnosis and Treatment of Venous Diseases
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.