Evaluation of an event-driven 3FI ASIC for spectroscopic X-ray detection with synchrotron radiation

The novel design and evaluation on the NSLS-II beamline of the 3FI application-specific integrated circuit (ASIC) bump-bonded to a simple, planar, 2D segmented silicon sensor are presented. The ASIC was developed for full-field fluorescence spectral X-ray imaging (3FI). It is a small-scale prototype that features a square array of 32 × 32 pixels, and the size of the pixels is 100 µm × 100 µm. The ASIC was implemented in a 65 nm CMOS integrated circuit fabrication process. Each pixel incorporates a charge-sensitive amplifier, a shaping filter, a discriminator, a peak detector and a sample-and-hold circuit, allowing detection of events and storage of signal amplitudes. The system operates in a frameless event-driven readout mode, outputting analog values for threshold-triggered events, allowing high-speed multi-element X-ray fluorescence data acquisition. The 3FI ASIC achieves per-channel spectrometric performance at a power consumption of only 200 µW per pixel, with nearly all dissipation confined to the analog front-end. An energy resolution is measured at the level of 308 eV full width at half-maximum (FWHM) at 8.04 keV (Cu Kα), and 138 eV FWHM at 3.69 keV (Ca Kα). This per-pixel capability makes the prototype suitable for in situ trace element microanalysis in biological and environmental studies. Moreover, the frameless architecture of the detector is designed to address limitations of conventional X-ray fluorescence microscopy, which typically requires mechanical scanning, by enabling continuous high-throughput data acquisition in future full-field implementations.

Authors

Institutions

Publication Details

Journal
Journal of Synchrotron Radiation
Published
2025-12-05
DOI
https://doi.org/10.1107/s1600577525010501
Primary Topic
Particle Detector Development and Performance
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of an event-driven 3FI ASIC for spectroscopic X-ray detection with synchrotron radiation

Abdul K. Rumaiz, G. Carini, Ryan Tappero, Tobias Kersten et al.
Journal of Synchrotron Radiation
Particle Detector Development and Performance
article

Evaluation of an event-driven 3FI ASIC for spectroscopic X-ray detection with synchrotron radiation

Abdul K. Rumaiz, G. Carini, Ryan Tappero, Tobias Kersten, D. Górni, A. Kuczewski, G. Deptuch, Donald A. Pinelli, Nicholas St John, Soumyajit Mandal, David P. Siddons, Piotr Maj, Giovanni Pinaroli
article en

Abstract

The novel design and evaluation on the NSLS-II beamline of the 3FI application-specific integrated circuit (ASIC) bump-bonded to a simple, planar, 2D segmented silicon sensor are presented. The ASIC was developed for full-field fluorescence spectral X-ray imaging (3FI). It is a small-scale prototype that features a square array of 32 × 32 pixels, and the size of the pixels is 100 µm × 100 µm. The ASIC was implemented in a 65 nm CMOS integrated circuit fabrication process. Each pixel incorporates a charge-sensitive amplifier, a shaping filter, a discriminator, a peak detector and a sample-and-hold circuit, allowing detection of events and storage of signal amplitudes. The system operates in a frameless event-driven readout mode, outputting analog values for threshold-triggered events, allowing high-speed multi-element X-ray fluorescence data acquisition. The 3FI ASIC achieves per-channel spectrometric performance at a power consumption of only 200 µW per pixel, with nearly all dissipation confined to the analog front-end. An energy resolution is measured at the level of 308 eV full width at half-maximum (FWHM) at 8.04 keV (Cu Kα), and 138 eV FWHM at 3.69 keV (Ca Kα). This per-pixel capability makes the prototype suitable for in situ trace element microanalysis in biological and environmental studies. Moreover, the frameless architecture of the detector is designed to address limitations of conventional X-ray fluorescence microscopy, which typically requires mechanical scanning, by enabling continuous high-throughput data acquisition in future full-field implementations.

Journal of Synchrotron RadiationVol. 33(1)
Brookhaven National Laboratory (US)
Reduced inequalities
Openalex Percentile: Top 23%
Particle Detector Development and Performance
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.