Quality control assessment and establishment of local diagnostic reference levels in two radiological centres in Osogbo, Nigeria

This study aims to assess Quality Control (QC) tests, determine the Entrance Surface Dose (ESD), and establish Local Diagnostic Reference Levels (LDRLs) for patients undergoing X-ray examinations at the University of Osun Teaching Hospital (UniOsun) and Union Diagnostic Centre (UDC), Osogbo, Nigeria. Demographic data from 334 patients, along with exposure parameters (kVp, mAs, FFD, and FSD), were recorded using a specially designed questionnaire with the assistance of radiographers. A one-time QC assessment test was conducted to verify the compliance of the X-ray machine with the international standard using a Gammex 330 Digital kV, dose, and time meter obtained from the National Institute for Radiation Protection and Research at the University of Ibadan, Nigeria. As referenced in the literature, a mathematical model was used to calculate the ESD (mGy) for various X-ray examinations conducted at UniOsun and UDC. Statistical analysis of the results was performed, and the 75th percentile for ESD was calculated. The results indicated that the mean highest ESD belonged to the Skull (0.047 ± 0.001) mGy and Lumbar spine (1.76 ± 0.48) mGy, and the lowest ESD belonged to the Elbow (0.012 ± 0.002) mGy and Elbow, Hand, Wrist (0.39 ± 0.06) mGy for UniOsun and UDC, respectively. The LDRL of the ESD ranged from (0.014-0.048) mGy to (0.42-2.29) mGy at UniOsun and UDC, respectively. In conclusion, the findings revealed a significant variation in ESD across different examinations and centres, highlighting the necessity for tailored dose optimisation strategies to ensure patient safety while maintaining diagnostic quality.

Authors

Institutions

Publication Details

Journal
Applied Radiation and Isotopes
Published
2025-12-22
DOI
https://doi.org/10.1016/j.apradiso.2025.112403
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Quality control assessment and establishment of local diagnostic reference levels in two radiological centres in Osogbo, Nigeria

C. J. Olowookere, Idowu Richard Akomolafe, U. E. Vincent, Naven Chetty et al.
Applied Radiation and Isotopes
Radiation Dose and Imaging
article

Quality control assessment and establishment of local diagnostic reference levels in two radiological centres in Osogbo, Nigeria

C. J. Olowookere, Idowu Richard Akomolafe, U. E. Vincent, Naven Chetty, Joel Olaniyi Onijingin, Jesutomisin Enoch Alegbeleye
article en

Abstract

This study aims to assess Quality Control (QC) tests, determine the Entrance Surface Dose (ESD), and establish Local Diagnostic Reference Levels (LDRLs) for patients undergoing X-ray examinations at the University of Osun Teaching Hospital (UniOsun) and Union Diagnostic Centre (UDC), Osogbo, Nigeria. Demographic data from 334 patients, along with exposure parameters (kVp, mAs, FFD, and FSD), were recorded using a specially designed questionnaire with the assistance of radiographers. A one-time QC assessment test was conducted to verify the compliance of the X-ray machine with the international standard using a Gammex 330 Digital kV, dose, and time meter obtained from the National Institute for Radiation Protection and Research at the University of Ibadan, Nigeria. As referenced in the literature, a mathematical model was used to calculate the ESD (mGy) for various X-ray examinations conducted at UniOsun and UDC. Statistical analysis of the results was performed, and the 75th percentile for ESD was calculated. The results indicated that the mean highest ESD belonged to the Skull (0.047 ± 0.001) mGy and Lumbar spine (1.76 ± 0.48) mGy, and the lowest ESD belonged to the Elbow (0.012 ± 0.002) mGy and Elbow, Hand, Wrist (0.39 ± 0.06) mGy for UniOsun and UDC, respectively. The LDRL of the ESD ranged from (0.014-0.048) mGy to (0.42-2.29) mGy at UniOsun and UDC, respectively. In conclusion, the findings revealed a significant variation in ESD across different examinations and centres, highlighting the necessity for tailored dose optimisation strategies to ensure patient safety while maintaining diagnostic quality.

Applied Radiation and IsotopesVol. 229
Redeemer's University (NG), University of Medical Sciences, Ondo, Lancaster University (GB), Olusegun Agagu University of Science and Technology (NG), University of KwaZulu-Natal (ZA)
Partnerships for the goals
Openalex Percentile: Top 32%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.