In Situ Dose Measurements in Brachytherapy Using Scintillation Detectors Based on the Al2O3:C, Al2O3:C,Mg, and GAGG:Ce Crystals

Currently, the use of scintillation crystals connected via optical fiber to a luminescence spectrometer (so-called fiber-optic dosimeters) offers a promising approach for real-time dosimetric measurements during brachytherapy treatments with γ-ray sources. This study aims to evaluate the applicability of fiber-optic dosimeters for in situ dose measurements during brachytherapy procedures, using Al<sub>2</sub>O<sub>3</sub>:C and Al<sub>2</sub>O<sub>3</sub>:C,Mg crystals, which have near-tissue density and effective atomic number (ρ = 3.99 g/cm<sup>3</sup>, Z<sub>eff</sub> = 10.8), as well as heavy GAGG:Ce scintillation crystals (ρ = 6.63 g/cm<sup>3</sup>, Z<sub>eff</sub> = 54.4). Radiation dose delivery was assessed through measurements of the resulting radioluminescence of the aforementioned scintillation crystals, connected via long optical fibers and recorded with highly sensitive, compact luminescence spectrometers. Measurements were performed in a dedicated phantom under clinical conditions at the Oncology Center in Bydgoszcz, Poland. The dosimeters were evaluated for in situ dose monitoring within the 0.5-8 Gy range during brachytherapy procedures using a <sup>192</sup>Ir (392 keV) source. The results showed a clear linear relationship between the delivered radiation dose and the scintillation output measured by the fiber-optic detector. The Gd<sub>3</sub>Al<sub>2.5</sub>Ga<sub>2.5</sub>O<sub>12</sub>:Ce crystal detector exhibited excellent linearity, while the Al<sub>2</sub>O<sub>3</sub>:C and Al<sub>2</sub>O<sub>3</sub>:C,Mg crystal detectors also showed a nearly linear dose-response relationship.

Authors

Institutions

Publication Details

Journal
Materials
Published
2025-12-22
DOI
https://doi.org/10.3390/ma19010045
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

In Situ Dose Measurements in Brachytherapy Using Scintillation Detectors Based on the Al2O3:C, Al2O3:C,Mg, and GAGG:Ce Crystals

Janusz Winiecki, Mark Akselrod, Bogna Sobiech, Sandra Witkiewicz-Lukaszek et al.
Materials
Advanced Radiotherapy Techniques
article

In Situ Dose Measurements in Brachytherapy Using Scintillation Detectors Based on the Al2O3:C, Al2O3:C,Mg, and GAGG:Ce Crystals

Janusz Winiecki, Mark Akselrod, Bogna Sobiech, Sandra Witkiewicz-Lukaszek, Yuriy Zorenko
article en

Abstract

Currently, the use of scintillation crystals connected via optical fiber to a luminescence spectrometer (so-called fiber-optic dosimeters) offers a promising approach for real-time dosimetric measurements during brachytherapy treatments with γ-ray sources. This study aims to evaluate the applicability of fiber-optic dosimeters for in situ dose measurements during brachytherapy procedures, using Al<sub>2</sub>O<sub>3</sub>:C and Al<sub>2</sub>O<sub>3</sub>:C,Mg crystals, which have near-tissue density and effective atomic number (ρ = 3.99 g/cm<sup>3</sup>, Z<sub>eff</sub> = 10.8), as well as heavy GAGG:Ce scintillation crystals (ρ = 6.63 g/cm<sup>3</sup>, Z<sub>eff</sub> = 54.4). Radiation dose delivery was assessed through measurements of the resulting radioluminescence of the aforementioned scintillation crystals, connected via long optical fibers and recorded with highly sensitive, compact luminescence spectrometers. Measurements were performed in a dedicated phantom under clinical conditions at the Oncology Center in Bydgoszcz, Poland. The dosimeters were evaluated for in situ dose monitoring within the 0.5-8 Gy range during brachytherapy procedures using a <sup>192</sup>Ir (392 keV) source. The results showed a clear linear relationship between the delivered radiation dose and the scintillation output measured by the fiber-optic detector. The Gd<sub>3</sub>Al<sub>2.5</sub>Ga<sub>2.5</sub>O<sub>12</sub>:Ce crystal detector exhibited excellent linearity, while the Al<sub>2</sub>O<sub>3</sub>:C and Al<sub>2</sub>O<sub>3</sub>:C,Mg crystal detectors also showed a nearly linear dose-response relationship.

MaterialsVol. 19(1)
Oklahoma State University (US), Kazimierz Wielki University in Bydgoszcz (PL), Collegium Medicum in Bydgoszcz (PL)
Openalex Percentile: Top 27%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

In Situ Dose Measurements in Brachytherapy Using Scintillation Detectors Based on the Al2O3:C, Al2O3:C,Mg, and GAGG:Ce Crystals — Janusz Winiecki, Mark Akselrod, et al. · Materials (2025) | TGRS Research Map | TGRS