Structural Stability of Silicone-Based Elastodontic Appliances After Clinical Use: Insights from FTIR Spectroscopy

<i>Background and Objectives:</i> Elastodontic appliances made of medical-grade silicone are increasingly used in interceptive orthodontics, but prolonged intraoral exposure may affect their stability. This study evaluated structural changes in LM-Activator<sup>TM</sup> 2 appliances after clinical use, using Fourier-transform infrared (FTIR) spectroscopy. <i>Materials and Methods:</i> Eight appliances (one unused control and seven worn for 3-24 months) were analyzed by FTIR-ATR in the 4000-650 cm<sup>-1</sup> range. Absorption bands characteristic of polydimethylsiloxane (PDMS) were quantified, and indices reflecting backbone crosslinking, side-group retention, hydrophilicity, and relative reduction in methyl-related spectral contributions were calculated. <i>Results:</i> The PDMS backbone remained chemically intact across all samples. However, progressive molecular reorganization was detected with wear duration. The Backbone Dominance Index increased significantly from control to 24 months, while side-group indices decreased, confirming apparent depletion of methyl-related FTIR bands. Hydrophilicity and crosslinking indices rose over time, particularly after 12 months, indicating increased surface polarity and network densification. <i>Conclusions:</i> LM-Activator<sup>TM</sup> 2 appliances undergo gradual intraoral aging, marked by backbone crosslinking and apparent reduction in methyl-associated vibrational contributions inferred from FTIR ratio side-groups. These changes, while not compromising the polymer identity, may influence surface properties, biofilm retention, and long-term mechanical behavior. Periodic replacement is recommended to ensure optimal clinical performance.

Authors

Institutions

Publication Details

Journal
Materials
Published
2025-12-19
DOI
https://doi.org/10.3390/ma19010013
Primary Topic
Dental materials and restorations
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Structural Stability of Silicone-Based Elastodontic Appliances After Clinical Use: Insights from FTIR Spectroscopy

Ştefania Dinu, Daniela Jumanca, Atena Gǎluşcan, Vanessa Bolchis et al.
Materials
Dental materials and restorations
article

Structural Stability of Silicone-Based Elastodontic Appliances After Clinical Use: Insights from FTIR Spectroscopy

Ştefania Dinu, Daniela Jumanca, Atena Gǎluşcan, Vanessa Bolchis, Emilia-Brindusa Brăilă, Vlase Titus, Vlase Gabriela, Vlad Tiberiu Alexa
article en

Abstract

<i>Background and Objectives:</i> Elastodontic appliances made of medical-grade silicone are increasingly used in interceptive orthodontics, but prolonged intraoral exposure may affect their stability. This study evaluated structural changes in LM-Activator<sup>TM</sup> 2 appliances after clinical use, using Fourier-transform infrared (FTIR) spectroscopy. <i>Materials and Methods:</i> Eight appliances (one unused control and seven worn for 3-24 months) were analyzed by FTIR-ATR in the 4000-650 cm<sup>-1</sup> range. Absorption bands characteristic of polydimethylsiloxane (PDMS) were quantified, and indices reflecting backbone crosslinking, side-group retention, hydrophilicity, and relative reduction in methyl-related spectral contributions were calculated. <i>Results:</i> The PDMS backbone remained chemically intact across all samples. However, progressive molecular reorganization was detected with wear duration. The Backbone Dominance Index increased significantly from control to 24 months, while side-group indices decreased, confirming apparent depletion of methyl-related FTIR bands. Hydrophilicity and crosslinking indices rose over time, particularly after 12 months, indicating increased surface polarity and network densification. <i>Conclusions:</i> LM-Activator<sup>TM</sup> 2 appliances undergo gradual intraoral aging, marked by backbone crosslinking and apparent reduction in methyl-associated vibrational contributions inferred from FTIR ratio side-groups. These changes, while not compromising the polymer identity, may influence surface properties, biofilm retention, and long-term mechanical behavior. Periodic replacement is recommended to ensure optimal clinical performance.

MaterialsVol. 19(1)
West University of Timişoara (RO), Victor Babeș University of Medicine and Pharmacy Timișoara (RO)
Openalex Percentile: Top 34%
Dental materials and restorations
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.