Practical Method for Evaluating the Element Sensitivity Variation of an Ultrasonic Annular Phased Array Transducer

The unique features of annular phased array transducers, such as ring-shaped elements and the concentric configuration, cause them to behave differently from commonly used linear array transducers, in terms of sound field distribution and pulse-echo response. Consequently, standard techniques for assessing linear array transducers can introduce significant errors when applied to annular array transducers, especially concerning element-to-element sensitivity variance. This study investigates the consistency of element sensitivity in annular phased array transducers. Through theoretical analysis, a Long-Belt source assumption model was developed based on the Rayleigh integral to characterize the responses of ring-shaped elements in an analytical and explicit form. The model suggests that the response amplitude is linearly correlated with the radial width of the element, which was validated by subsequent numerical simulations. Based on these findings, a modified sensitivity evaluation algorithm for annular array transducers is presented. The response voltage per unit width, rather than the total response voltage, is used to eliminate the influence of varying geometries and sizes across elements. The sensitivity variation of a 32-element annular array transducer was evaluated using the new algorithm. Compared to the uncorrected measurement, the maximum sensitivity variation was reduced significantly from 25 dB to 6 dB, revealing the transducer's intrinsic consistency despite the different geometric features of each element. Due to its distinct geometry compared to the ring-shaped elements, the central element cannot be corrected or evaluated using this method. These results suggest that the proposed algorithm enables the more accurate evaluation of sensitivity consistency for annular phased array transducers, thereby improving measurement reliability in practical applications.

Authors

Institutions

Publication Details

Journal
Sensors
Published
2025-12-19
DOI
https://doi.org/10.3390/s26010025
Primary Topic
Ultrasound Imaging and Elastography
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Practical Method for Evaluating the Element Sensitivity Variation of an Ultrasonic Annular Phased Array Transducer

Xiao Guang Liu, XT Zhou, Zhengxiao Sha, Yanze Liu et al.
Sensors
Ultrasound Imaging and Elastography
article

Practical Method for Evaluating the Element Sensitivity Variation of an Ultrasonic Annular Phased Array Transducer

Xiao Guang Liu, XT Zhou, Zhengxiao Sha, Yanze Liu, Xiao Wang
article en

Abstract

The unique features of annular phased array transducers, such as ring-shaped elements and the concentric configuration, cause them to behave differently from commonly used linear array transducers, in terms of sound field distribution and pulse-echo response. Consequently, standard techniques for assessing linear array transducers can introduce significant errors when applied to annular array transducers, especially concerning element-to-element sensitivity variance. This study investigates the consistency of element sensitivity in annular phased array transducers. Through theoretical analysis, a Long-Belt source assumption model was developed based on the Rayleigh integral to characterize the responses of ring-shaped elements in an analytical and explicit form. The model suggests that the response amplitude is linearly correlated with the radial width of the element, which was validated by subsequent numerical simulations. Based on these findings, a modified sensitivity evaluation algorithm for annular array transducers is presented. The response voltage per unit width, rather than the total response voltage, is used to eliminate the influence of varying geometries and sizes across elements. The sensitivity variation of a 32-element annular array transducer was evaluated using the new algorithm. Compared to the uncorrected measurement, the maximum sensitivity variation was reduced significantly from 25 dB to 6 dB, revealing the transducer's intrinsic consistency despite the different geometric features of each element. Due to its distinct geometry compared to the ring-shaped elements, the central element cannot be corrected or evaluated using this method. These results suggest that the proposed algorithm enables the more accurate evaluation of sensitivity consistency for annular phased array transducers, thereby improving measurement reliability in practical applications.

SensorsVol. 26(1)
Beijing Institute of Aeronautical Materials (CN), Aero Engine Corporation of China (China) (CN)
Openalex Percentile: Top 32%
Ultrasound Imaging and Elastography
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Practical Method for Evaluating the Element Sensitivity Variation of an Ultrasonic Annular Phased Array Transducer — Xiao Guang Liu, XT Zhou, et al. · Sensors (2025) | TGRS Research Map | TGRS