Mechanical and Biological Behaviors of Body Centered Cubic and Cubic Lattices for Hip Implant Applications

A suitable combination of mechanical, topographical, and biological responses enhances implant‐bone interaction and ensures successful patient outcomes. In this regard, body‐centered‐cubic (BCC) and cubic lattices are designed and fabricated with porosity ranging between 50% and 80%. Finite element analysis and uniaxial tensile tests are conducted to evaluate the mechanical properties of the lattices. Roughness and wettability are studied to evaluate the topographical properties. The biological behavior is studied through in vitro response of MG‐63 cells on scaffolds after 1, 4, and 7 d. Two types of surfaces, as‐fabricated and heat‐treated, are prepared for roughness, wettability measurements, and in vitro experiments. The mechanical tests reveal that BCC and cubic lattices can mimic femoral bone properties at 60% and 65% porosity, respectively. The R a values are found to be ≈12 μm for the as‐produced and 1.2 to 3 μm for the heat‐treated lattices. Hydrophobic and super‐hydrophilic natures have been observed with as‐fabricated and heat‐treated scaffolds during the wettability experiment, respectively. The in vitro response shows the most favorable cell proliferations in heat‐treated cubic lattices compared to other scaffolds. Based on these observations, a novel porous short hip stem featuring a cubic porous structure has been designed and successfully fabricated using selective laser melting.

Authors

Institutions

Publication Details

Journal
Advanced Engineering Materials
Published
2025-12-21
DOI
https://doi.org/10.1002/adem.202502326
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Mechanical and Biological Behaviors of Body Centered Cubic and Cubic Lattices for Hip Implant Applications

Prashanta K. Mahato, Bidyut Pal, Abhik Chaudhuri, Sarthak S. Singh
Advanced Engineering Materials
Orthopaedic implants and arthroplasty
article

Mechanical and Biological Behaviors of Body Centered Cubic and Cubic Lattices for Hip Implant Applications

Prashanta K. Mahato, Bidyut Pal, Abhik Chaudhuri, Sarthak S. Singh
article en

Abstract

A suitable combination of mechanical, topographical, and biological responses enhances implant‐bone interaction and ensures successful patient outcomes. In this regard, body‐centered‐cubic (BCC) and cubic lattices are designed and fabricated with porosity ranging between 50% and 80%. Finite element analysis and uniaxial tensile tests are conducted to evaluate the mechanical properties of the lattices. Roughness and wettability are studied to evaluate the topographical properties. The biological behavior is studied through in vitro response of MG‐63 cells on scaffolds after 1, 4, and 7 d. Two types of surfaces, as‐fabricated and heat‐treated, are prepared for roughness, wettability measurements, and in vitro experiments. The mechanical tests reveal that BCC and cubic lattices can mimic femoral bone properties at 60% and 65% porosity, respectively. The R a values are found to be ≈12 μm for the as‐produced and 1.2 to 3 μm for the heat‐treated lattices. Hydrophobic and super‐hydrophilic natures have been observed with as‐fabricated and heat‐treated scaffolds during the wettability experiment, respectively. The in vitro response shows the most favorable cell proliferations in heat‐treated cubic lattices compared to other scaffolds. Based on these observations, a novel porous short hip stem featuring a cubic porous structure has been designed and successfully fabricated using selective laser melting.

Advanced Engineering MaterialsVol. 28(3)
Indian Institute of Technology Dhanbad (IN), Indian Institute of Engineering Science and Technology, Shibpur (IN)
Openalex Percentile: Top 28%
Orthopaedic implants and arthroplasty
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Mechanical and Biological Behaviors of Body Centered Cubic and Cubic Lattices for Hip Implant Applications — Prashanta K. Mahato, Bidyut Pal, et al. · Advanced Engineering Materials (2025) | TGRS Research Map | TGRS