Application of a CdTe Photovoltaic Dosimeter to Therapeutic Megavoltage Photon Beams
Accurate real-time dosimetry is key in megavoltage radiotherapy; however, many detectors require external biasing or complex instrumentation. This study evaluated thin-film CdTe solar cells operating in photovoltaic (zero-bias) mode as medical dosimeters. Superstrate ITO/CdS/CdTe/Cu/Au devices were fabricated and irradiated with 6-MV photons from a clinical linear accelerator to 20 kGy cumulative dose. Electrical and dosimetric properties were assessed based on AM 1.5 current–voltage measurements, external quantum efficiency (EQE), dose linearity, dose-rate dependence, field-size dependence, percentage depth dose (PDD), and one-month reproducibility. With increasing dose (5–20 kGy), the open-circuit voltage and fill factor decreased by ~2–3%, the short-circuit current density by ~10%, retaining ~87% initial efficiency. Series and shunt resistances were stable, while EQE decreased uniformly (~5%), indicating degradation mainly from increased nonradiative recombination. Dose–signal linearity remained intact, and post-irradiation sensitivity loss was corrected with a single calibration factor. Dose-rate dependence was minor; low reverse bias (~3–7 V) enhanced response without nonlinearity. Field-size and PDD responses agreed with ionization chamber data within ~1%, and weekly stability was within ~1%. Parallel stacking of two cells increased signal nearly linearly. CdTe solar-cell detectors thus enable zero-bias, real-time, stable, and scalable dosimetry and strongly agree with reference standards.
Institutions
- Chonnam National University (KR)
- Korea University (KR)
- The Catholic University of Korea St. Vincent's Hospital (KR)
- Jeonbuk National University Hospital (KR)
- Jeonbuk National University (KR)
- Catholic University of Korea (KR)
Publication Details
- Journal
- Applied Sciences
- Published
- 2025-12-12
- DOI
- https://doi.org/10.3390/app152413091
- Primary Topic
- Advanced Radiotherapy Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- Mike and Josie Harper Cancer Research Institute
- National Research Foundation
- Chonnam National University
- National Research Foundation of Korea
- Ministry of Science and ICT, South Korea