Robust Torque Control in Flexion/ Extension Wrist Movement

This study explores the feasibility of implementing robust torque control in a wrist flexion and extension rehabilitation device using a proportional-integral-controller (PID) controller. A simulation model was developed, consisting of an ideal DC motor, a biomechanical hand model, a torque estimator, and a PID feedback controller. The wrist was modeled as a rotational mass-spring-damper system, capturing essential mechanical behavior while enabling rapid prototyping. The control system used a sigmoid torque input to mimic gradual clinical force application and avoid abrupt joint loading. Robustness was evaluated by varying biomechanical parameters across 1,000 simulations. Results showed consistent performance, with maximum overshoot below 3% and settling times under 0.22 seconds. The controller also maintained accurate torque tracking under external disturbances. These findings indicate that a well-tuned PID controller can provide reliable torque control despite variability in human biomechanics.

Institutions

Publication Details

Published
2025-09-03
DOI
https://doi.org/10.1145/3748699.3749806
Primary Topic
Stroke Rehabilitation and Recovery
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Robust Torque Control in Flexion/ Extension Wrist Movement

Stroke Rehabilitation and Recovery
article

Robust Torque Control in Flexion/ Extension Wrist Movement

article en

Abstract

This study explores the feasibility of implementing robust torque control in a wrist flexion and extension rehabilitation device using a proportional-integral-controller (PID) controller. A simulation model was developed, consisting of an ideal DC motor, a biomechanical hand model, a torque estimator, and a PID feedback controller. The wrist was modeled as a rotational mass-spring-damper system, capturing essential mechanical behavior while enabling rapid prototyping. The control system used a sigmoid torque input to mimic gradual clinical force application and avoid abrupt joint loading. Robustness was evaluated by varying biomechanical parameters across 1,000 simulations. Results showed consistent performance, with maximum overshoot below 3% and settling times under 0.22 seconds. The controller also maintained accurate torque tracking under external disturbances. These findings indicate that a well-tuned PID controller can provide reliable torque control despite variability in human biomechanics.

University of Antwerp (BE), Ghent University (BE)
Openalex Percentile: Top 36%
Stroke Rehabilitation and Recovery
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.