S2170 Currently Available Guidewire Locking Device (GLD) Designs in the Market Can Deform Expensive Endoscopic Retrograde Cholangiopancreatography (ERCP) Accessories and Compromise the Integrity of Guidewires

Introduction: GLDs used in ERCP secure guidewires and prevent fluid reflux through the biopsy channel. However, their internal design may create resistance during accessory insertion, and overly tight seals can deform devices, impacting procedural success. Repeated locking–unlocking cycles, especially in the presence of bile and contrast, may compromise the polytetrafluoroethylene (PTFE) coating on guidewires. This study aimed to evaluate the impact of current GLD designs on ERCP accessory integrity and guidewire durability. Methods: Three commercial GLDs were tested using a validated Newton gauge. We measured the force required to permanently deform 3 ERCP accessories—sphincterotome, stone-extraction balloon, and dilating balloon—and the insertion force for each (10 insertions/accessory/GLD). PTFE integrity of guidewires exposed to bile and contrast was evaluated after ≥15 lock–unlock cycles. Damage was categorized as holes, fraying, or linear fractures, and scored 1–3 (1 = none, 3 = severe). ANOVA was used for comparisons. Deformation was also assessed as a binary variable (above vs below threshold), with odds ratios calculated. Results: The force required to permanently deform the sphincterotome was 4.70 N, stone-extraction balloon 1.60 N, and dilating balloon 0.88 N. Insertion forces differed significantly across GLDs for each accessory. For sphincterotomes, mean forces were: GLD#1 = 5.21 ± 0.94 N, GLD#2 = 5.41 ± 0.35 N, and GLD#3 = 2.47 ± 0.22 N (F = 62.1, P < 0.001); only GLD#3 did not cause deformation. For stone-extraction balloons: GLD#1 = 6.16 ± 0.22 N, GLD#2 = 11.49 ± 0.72 N, GLD#3 = 2.62 ± 0.25 N (F = 970.3, P < 0.001); again, only GLD#3 remained below the deformation threshold. For CRE balloons: GLD#1 = 5.92 ± 0.20 N, GLD#2 = 5.54 ± 0.30 N, GLD#3 = 2.68 ± 0.20 N; however, all 3 exceeded the deformation threshold. When analyzed as a binary outcome, GLD#3 had zero deformation events in sphincterotome and stone balloon testing, while GLD#1 and GLD#2 had 100%, yielding significant odds ratios favoring GLD#3 (P < 0.005 for all comparisons). PTFE integrity analysis showed severe damage with GLD#2 (score: 8/9; holes and fractures), moderate with GLD#1 (5/9), and mild with GLD#3 (3/9). Conclusion: GLD designs vary in insertion resistance, with some exceeding deformation thresholds for ERCP devices. Repeated lock–unlock cycles, particularly with bile and contrast, can damage guidewire PTFE coating. Design refinements are needed to minimize device damage, cost, and risk.

Publication Details

Journal
The American Journal of Gastroenterology
Published
2025-10-01
DOI
https://doi.org/10.14309/01.ajg.0001136140.52826.11
Primary Topic
Gallbladder and Bile Duct Disorders
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

S2170 Currently Available Guidewire Locking Device (GLD) Designs in the Market Can Deform Expensive Endoscopic Retrograde Cholangiopancreatography (ERCP) Accessories and Compromise the Integrity of Guidewires

The American Journal of Gastroenterology
Gallbladder and Bile Duct Disorders
article

S2170 Currently Available Guidewire Locking Device (GLD) Designs in the Market Can Deform Expensive Endoscopic Retrograde Cholangiopancreatography (ERCP) Accessories and Compromise the Integrity of Guidewires

article en

Abstract

Introduction: GLDs used in ERCP secure guidewires and prevent fluid reflux through the biopsy channel. However, their internal design may create resistance during accessory insertion, and overly tight seals can deform devices, impacting procedural success. Repeated locking–unlocking cycles, especially in the presence of bile and contrast, may compromise the polytetrafluoroethylene (PTFE) coating on guidewires. This study aimed to evaluate the impact of current GLD designs on ERCP accessory integrity and guidewire durability. Methods: Three commercial GLDs were tested using a validated Newton gauge. We measured the force required to permanently deform 3 ERCP accessories—sphincterotome, stone-extraction balloon, and dilating balloon—and the insertion force for each (10 insertions/accessory/GLD). PTFE integrity of guidewires exposed to bile and contrast was evaluated after ≥15 lock–unlock cycles. Damage was categorized as holes, fraying, or linear fractures, and scored 1–3 (1 = none, 3 = severe). ANOVA was used for comparisons. Deformation was also assessed as a binary variable (above vs below threshold), with odds ratios calculated. Results: The force required to permanently deform the sphincterotome was 4.70 N, stone-extraction balloon 1.60 N, and dilating balloon 0.88 N. Insertion forces differed significantly across GLDs for each accessory. For sphincterotomes, mean forces were: GLD#1 = 5.21 ± 0.94 N, GLD#2 = 5.41 ± 0.35 N, and GLD#3 = 2.47 ± 0.22 N (F = 62.1, P < 0.001); only GLD#3 did not cause deformation. For stone-extraction balloons: GLD#1 = 6.16 ± 0.22 N, GLD#2 = 11.49 ± 0.72 N, GLD#3 = 2.62 ± 0.25 N (F = 970.3, P < 0.001); again, only GLD#3 remained below the deformation threshold. For CRE balloons: GLD#1 = 5.92 ± 0.20 N, GLD#2 = 5.54 ± 0.30 N, GLD#3 = 2.68 ± 0.20 N; however, all 3 exceeded the deformation threshold. When analyzed as a binary outcome, GLD#3 had zero deformation events in sphincterotome and stone balloon testing, while GLD#1 and GLD#2 had 100%, yielding significant odds ratios favoring GLD#3 (P < 0.005 for all comparisons). PTFE integrity analysis showed severe damage with GLD#2 (score: 8/9; holes and fractures), moderate with GLD#1 (5/9), and mild with GLD#3 (3/9). Conclusion: GLD designs vary in insertion resistance, with some exceeding deformation thresholds for ERCP devices. Repeated lock–unlock cycles, particularly with bile and contrast, can damage guidewire PTFE coating. Design refinements are needed to minimize device damage, cost, and risk.

The American Journal of GastroenterologyVol. 120(10S2)
Openalex Percentile: Top 30%
Gallbladder and Bile Duct Disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.