Comparison of Ice Lined Refrigerators of the different types used in Western part of India under the Universal Immunization Program

Background & Objective: The effectiveness of Ice-Lined Refrigerators (ILRs) is crucial in maintaining the cold chain for vaccines, particularly in India's Universal Immunization Program (UIP). This cross-sectional study aimed to compare the performance of top-and front-opening ILRs in Gujarat, focusing on temperature stability, holdover time, and functionality. Methods: A total of 123 ILRs from various manufacturers were assessed across urban and rural health facilities. Key parameters such as ambient temperature, breakdown frequency, ease of use, and temperature fluctuations during door openings were recorded. Mann-whitney U test/ t-test and Binary logistic regression model were used as statistical methods. Results: Analysis indicated that top-opening ILRs, which made up 72.4% of the sample, performed significantly better in maintaining temperature stability and had a longer holdover time (mean: 5.4hours) compared to front-opening ILRs (mean: 4.3hours). Temperature breaches were more frequent in top-opening models (34.8%) compared to front-opening ones (5.9%), but the front-opening ILRs exhibited a greater temperature rise during door openings. Logistic regression analysis revealed a strong association between longer holdover time and top-opening ILRs. The study also found infrastructure and training gaps at some facilities, with only 57.7% of ILRs equipped with functional temperature monitoring systems, and many health workers lacking updated training. The findings suggest that top-opening ILRs are more reliable for cold chain management, though further investigation into front-opening models is warranted. Conclusion: For future cold chain management, prioritize top-opening ILRs for their reliability, increase AMC coverage for consistent maintenance, and enhance training for personnel on ILR handling, especially with newer models. Additionally, invest in backup-power and temperature monitoring equipment to prevent temperature breaches, particularly in remote areas.

Publication Details

Journal
Indian Journal of Community Health (Indian Association of Preventive and Social Medicine)
Published
2025-08-31
Primary Topic
Infection Control and Ventilation
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparison of Ice Lined Refrigerators of the different types used in Western part of India under the Universal Immunization Program

Indian Journal of Community Health (Indian Association of Preventive and Social Medicine)
Infection Control and Ventilation
article

Comparison of Ice Lined Refrigerators of the different types used in Western part of India under the Universal Immunization Program

article en

Abstract

Background & Objective: The effectiveness of Ice-Lined Refrigerators (ILRs) is crucial in maintaining the cold chain for vaccines, particularly in India's Universal Immunization Program (UIP). This cross-sectional study aimed to compare the performance of top-and front-opening ILRs in Gujarat, focusing on temperature stability, holdover time, and functionality. Methods: A total of 123 ILRs from various manufacturers were assessed across urban and rural health facilities. Key parameters such as ambient temperature, breakdown frequency, ease of use, and temperature fluctuations during door openings were recorded. Mann-whitney U test/ t-test and Binary logistic regression model were used as statistical methods. Results: Analysis indicated that top-opening ILRs, which made up 72.4% of the sample, performed significantly better in maintaining temperature stability and had a longer holdover time (mean: 5.4hours) compared to front-opening ILRs (mean: 4.3hours). Temperature breaches were more frequent in top-opening models (34.8%) compared to front-opening ones (5.9%), but the front-opening ILRs exhibited a greater temperature rise during door openings. Logistic regression analysis revealed a strong association between longer holdover time and top-opening ILRs. The study also found infrastructure and training gaps at some facilities, with only 57.7% of ILRs equipped with functional temperature monitoring systems, and many health workers lacking updated training. The findings suggest that top-opening ILRs are more reliable for cold chain management, though further investigation into front-opening models is warranted. Conclusion: For future cold chain management, prioritize top-opening ILRs for their reliability, increase AMC coverage for consistent maintenance, and enhance training for personnel on ILR handling, especially with newer models. Additionally, invest in backup-power and temperature monitoring equipment to prevent temperature breaches, particularly in remote areas.

Indian Journal of Community Health (Indian Association of Preventive and Social Medicine)
Industry, innovation and infrastructure
Openalex Percentile: Top 31%
Infection Control and Ventilation
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.