Combined leadless pacing and subcutaneous defibrillation strategy in a high-risk patient: first case report from Peru
We present the case of a 51-year-old male with non-ischemic dilated cardiomyopathy and complete atrioventricular block, who was previously implanted with a cardiac resynchronization therapy defibrillator. The patient developed signs of pocket infection with a high risk of extrusion. Partial system extraction was performed, followed by 14 days of intravenous antibiotic therapy. Due to a history of ventricular fibrillation and permanent pacing dependency, and in the absence of viable transvenous access, a sequential implantation strategy was adopted using a leadless pacemaker (Micra AV, Medtronic) and a subcutaneous implantable cardioverter-defibrillator (EMBLEM, Boston Scientific). Both procedures were completed without complications, and the patient showed favorable recovery, with effective pacing, no arrhythmic recurrences, and no signs of infection at the six-month follow-up. This case illustrates the feasibility of a fully leadless approach in high-risk patients with contraindications to conventional transvenous systems.
Institutions
- The Arrhythmia Institute (US)
- EsSALUD (PE)
Publication Details
- Journal
- Archivos Peruanos de Cardiología y Cirugía Cardiovascular
- Published
- 2025-10-18
- DOI
- https://doi.org/10.47487/apcyccv.v6i4.513
- Primary Topic
- Cardiac pacing and defibrillation studies
- Type
- article
- Field-Weighted Citation Impact
- 0.00