Objective Evaluation and Comparison of Diffractive versus Refractive Phakic Intraocular Lenses for Presbyopia Correction

Agnieszka Zielinska,1,2 Ava Niknahad,1 Grzegorz Labuz,1 Zhiyi Wu,1,3 Hyeck-Soo Son,1 Gerd U Auffarth,1 Ramin Khoramnia1,4 1The David J. Apple Center for Vision Research, Department of Ophthalmology, University of Heidelberg, Heidelberg, Germany; 2Institute of Physics, Faculty of Physics, Astronomy and Informatics, Nicolaus Copernicus University in Torun, Torun, Poland; 3Eye Center, The Second Affiliated Hospital, School of Medicine, Zhejiang University, Hangzhou, People’s Republic of China; 4Department of Ophthalmology, University Hospital Carl Gustav Carus, TU Dresden, Dresden, GermanyCorrespondence: Grzegorz Labuz, The David J. Apple Center for Vision Research, Department of Ophthalmology, University of Heidelberg, Im Neuenheimer Feld 400, Heidelberg, 69120, Germany, Tel +49 6221 56 6999, Fax +49 6221 56 5422, Email [email protected]: Phakic intraocular lenses (IOLs) have newly been introduced for presbyopia correction, with no studies to date directly comparing the various models to one another. This study assesses and compares the optical quality of two phakic IOLs - a diffractive IPCL (Care Group) with +2.5D added power and a refractive Artiplus (Ophtec) - in a laboratory setting.Methods: A custom-made insert contained a − 3D phakic IOL and a 23D monofocal (Precizon Monofocal, Ophtec). Two samples of each phakic IOL were tested using OptiSpheric IOL PRO2 under spherical and chromatic-aberration conditions at 3 and 4.5mm apertures at near (40cm), intermediate (67cm) and far distances. Modulation transfer function (MTF) related variables, and the 1951 USAF resolution target images were compared.Results: The MTFs at far distances, 67cm, and 40cm revealed minor differences between the two IOLs for a 3-mm pupil. Simulated visual acuity showed no disparity at distance but improved by 0.01 and 0.05 logMAR with Artiplus at intermediate and near points. Additionally, Artiplus exhibited a plateau in intermediate performance, while the IPCL demonstrated higher optical quality close to 50cm, corresponding to its secondary-focus location. Both lenses showed reduced optical quality with larger apertures. The USAF resolution target results supported these findings.Conclusion: Artiplus and IPCL (+2.5D add) showed good optical quality across the focus range, with Artiplus having a flatter simulated defocus curve. While the expected visual acuity differences were minimal, they may become more pronounced in patients with larger pupils, warranting further investigation.Keywords: Artiplus, IPCL, Phakic IOLs, Defocus Curve, Presbyopia correction

Institutions

Publication Details

Journal
SHILAP Revista de lepidopterología
Published
2025-10-01
Primary Topic
Ophthalmology and Visual Impairment Studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Objective Evaluation and Comparison of Diffractive versus Refractive Phakic Intraocular Lenses for Presbyopia Correction

SHILAP Revista de lepidopterología
Ophthalmology and Visual Impairment Studies
article

Objective Evaluation and Comparison of Diffractive versus Refractive Phakic Intraocular Lenses for Presbyopia Correction

article en

Abstract

Agnieszka Zielinska,1,2 Ava Niknahad,1 Grzegorz Labuz,1 Zhiyi Wu,1,3 Hyeck-Soo Son,1 Gerd U Auffarth,1 Ramin Khoramnia1,4 1The David J. Apple Center for Vision Research, Department of Ophthalmology, University of Heidelberg, Heidelberg, Germany; 2Institute of Physics, Faculty of Physics, Astronomy and Informatics, Nicolaus Copernicus University in Torun, Torun, Poland; 3Eye Center, The Second Affiliated Hospital, School of Medicine, Zhejiang University, Hangzhou, People’s Republic of China; 4Department of Ophthalmology, University Hospital Carl Gustav Carus, TU Dresden, Dresden, GermanyCorrespondence: Grzegorz Labuz, The David J. Apple Center for Vision Research, Department of Ophthalmology, University of Heidelberg, Im Neuenheimer Feld 400, Heidelberg, 69120, Germany, Tel +49 6221 56 6999, Fax +49 6221 56 5422, Email [email protected]: Phakic intraocular lenses (IOLs) have newly been introduced for presbyopia correction, with no studies to date directly comparing the various models to one another. This study assesses and compares the optical quality of two phakic IOLs - a diffractive IPCL (Care Group) with +2.5D added power and a refractive Artiplus (Ophtec) - in a laboratory setting.Methods: A custom-made insert contained a − 3D phakic IOL and a 23D monofocal (Precizon Monofocal, Ophtec). Two samples of each phakic IOL were tested using OptiSpheric IOL PRO2 under spherical and chromatic-aberration conditions at 3 and 4.5mm apertures at near (40cm), intermediate (67cm) and far distances. Modulation transfer function (MTF) related variables, and the 1951 USAF resolution target images were compared.Results: The MTFs at far distances, 67cm, and 40cm revealed minor differences between the two IOLs for a 3-mm pupil. Simulated visual acuity showed no disparity at distance but improved by 0.01 and 0.05 logMAR with Artiplus at intermediate and near points. Additionally, Artiplus exhibited a plateau in intermediate performance, while the IPCL demonstrated higher optical quality close to 50cm, corresponding to its secondary-focus location. Both lenses showed reduced optical quality with larger apertures. The USAF resolution target results supported these findings.Conclusion: Artiplus and IPCL (+2.5D add) showed good optical quality across the focus range, with Artiplus having a flatter simulated defocus curve. While the expected visual acuity differences were minimal, they may become more pronounced in patients with larger pupils, warranting further investigation.Keywords: Artiplus, IPCL, Phakic IOLs, Defocus Curve, Presbyopia correction

SHILAP Revista de lepidopterología
Apple (Israel) (IL), Ophthalmology Associates (United States) (US)
Openalex Percentile: Top 37%
Ophthalmology and Visual Impairment Studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.