Enhancing Image Quality While Reducing Dose: A Phantom Study on 100 kVp CT Protocol with iDose

This study evaluates how effectively an optimized CT imaging technique can lower patient radiation exposure while preserving approximate diagnostic image quality using a standardized phantom model (ACR CT). On a Philips Access CT scanner, two distinct scanning configurations were used: an optimized method employing 100 kVp in conjunction with sophisticated iterative image reconstruction, and a traditional method using 120 kVp and filtered back projection (FBP). The American College of Radiology (ACR) phantom was the main model used to evaluate important imaging characteristics such signal uniformity, contrast differentiation, and spatial-definition. The volume CT dose index (CTDIvol), dose-length product (DLP), and estimated effective dose were among the dose indications used to evaluate radiation exposure. Every acquisition was completed in duplicate to ensure uniformity. the data was analytically processed using SPSS software (version 27), and findings with p-values less than 0.05 were considered statistically significant in addition to In comparison to the standard protocol, the optimized low-dose protocol (100 kVp, iDose⁴) showed a considerable reduction in radiation dosage, with CTDIvol falling by 29.8% and the effective dose by 27.4% (p < 0.001). There were no statistically significant variations in the procedures' spatial resolution, homogeneity, or CT number accuracy, even with the low dosage settings. Noise reduction and CNR were finally greatly enhanced using the iDose reconstruction technique. The results validate that a low-dose protocol that is tuned and uses iDose⁴ reconstruction reduces radiation dose significantly while maintaining picture quality that is clinically acceptable. Its application in clinical practice to improve patient safety and operational efficiency is supported by phantom-based validation

Institutions

Publication Details

Journal
Zenodo (CERN European Organization for Nuclear Research)
Published
2025-09-25
DOI
https://doi.org/10.5281/zenodo.17342950
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Enhancing Image Quality While Reducing Dose: A Phantom Study on 100 kVp CT Protocol with iDose

Zenodo (CERN European Organization for Nuclear Research)
Radiation Dose and Imaging
article

Enhancing Image Quality While Reducing Dose: A Phantom Study on 100 kVp CT Protocol with iDose

article en

Abstract

This study evaluates how effectively an optimized CT imaging technique can lower patient radiation exposure while preserving approximate diagnostic image quality using a standardized phantom model (ACR CT). On a Philips Access CT scanner, two distinct scanning configurations were used: an optimized method employing 100 kVp in conjunction with sophisticated iterative image reconstruction, and a traditional method using 120 kVp and filtered back projection (FBP). The American College of Radiology (ACR) phantom was the main model used to evaluate important imaging characteristics such signal uniformity, contrast differentiation, and spatial-definition. The volume CT dose index (CTDIvol), dose-length product (DLP), and estimated effective dose were among the dose indications used to evaluate radiation exposure. Every acquisition was completed in duplicate to ensure uniformity. the data was analytically processed using SPSS software (version 27), and findings with p-values less than 0.05 were considered statistically significant in addition to In comparison to the standard protocol, the optimized low-dose protocol (100 kVp, iDose⁴) showed a considerable reduction in radiation dosage, with CTDIvol falling by 29.8% and the effective dose by 27.4% (p < 0.001). There were no statistically significant variations in the procedures' spatial resolution, homogeneity, or CT number accuracy, even with the low dosage settings. Noise reduction and CNR were finally greatly enhanced using the iDose reconstruction technique. The results validate that a low-dose protocol that is tuned and uses iDose⁴ reconstruction reduces radiation dose significantly while maintaining picture quality that is clinically acceptable. Its application in clinical practice to improve patient safety and operational efficiency is supported by phantom-based validation

Zenodo (CERN European Organization for Nuclear Research)
Middle Technical University (IQ)
Good health and well-being
Openalex Percentile: Top 36%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.