CALIBRATION THERMOCOUPLE K-TYPE ON THE RANGE TEMPERATURE -10°C TO 30°C USING ASTM THERMOMETER AS REFERENCE

This research focuses on the calibration of four type K thermocouples in the temperature range -10°C to 30°C, using chamber water bath with high precision. The calibration process is carried out by comparing thermocouple readings with readings from ASTM standard thermometers, with data acquisition managed using National Instruments hardware and LabVIEW software. The results of research on the accuracy of type K thermocouples show a shift in linearity and uncertainty values/ uncertainty are within acceptable engineering tolerances. A linear correction equation for each thermocouple is derived to minimize measurement error, ensuring that the thermocouple provides reliable temperature readings for applications within the specified range. This research shows the importance of using high-precision calibration systems, such as ASTM thermometers, to improve thermocouple measurement accuracy, providing a comprehensive solution for precision temperature measurements.

Institutions

Publication Details

Journal
Journal of Multidisciplinary Research and Technology
Published
2025-01-09
DOI
https://doi.org/10.63891/j-mart.v1i1.6
Primary Topic
Agriculture, Water, and Health
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

CALIBRATION THERMOCOUPLE K-TYPE ON THE RANGE TEMPERATURE -10°C TO 30°C USING ASTM THERMOMETER AS REFERENCE

Journal of Multidisciplinary Research and Technology
Agriculture, Water, and Health
article

CALIBRATION THERMOCOUPLE K-TYPE ON THE RANGE TEMPERATURE -10°C TO 30°C USING ASTM THERMOMETER AS REFERENCE

article en

Abstract

This research focuses on the calibration of four type K thermocouples in the temperature range -10°C to 30°C, using chamber water bath with high precision. The calibration process is carried out by comparing thermocouple readings with readings from ASTM standard thermometers, with data acquisition managed using National Instruments hardware and LabVIEW software. The results of research on the accuracy of type K thermocouples show a shift in linearity and uncertainty values/ uncertainty are within acceptable engineering tolerances. A linear correction equation for each thermocouple is derived to minimize measurement error, ensuring that the thermocouple provides reliable temperature readings for applications within the specified range. This research shows the importance of using high-precision calibration systems, such as ASTM thermometers, to improve thermocouple measurement accuracy, providing a comprehensive solution for precision temperature measurements.

Journal of Multidisciplinary Research and TechnologyVol. 1(1)
Gunadarma University (ID)
Openalex Percentile: Top 46%
Agriculture, Water, and Health
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.