High-count-rate effects in event processing for the XRISM/Resolve X-ray microcalorimeter. II. Energy scale and resolution in orbit

The Resolve instrument on the X-ray Imaging and Spectroscopy Mission (XRISM) uses a 36 pixel microcalorimeter designed to deliver high-resolution, non-dispersive X-ray spectroscopy. Although it is optimized for extended sources with low count rates, Resolve observations of bright point sources are still able to provide unique insights into the physics of these objects, as long as high-count-rate effects are addressed in the analysis. These effects include the loss of exposure time for each pixel, changes in the energy scale, and changes in the energy resolution. To investigate these effects under realistic observational conditions, we observed the bright X-ray source, the Crab Nebula, with XRISM at several offset positions with respect to the Resolve field of view and with continuous illumination from 55Fe sources on the filter wheel. For the spectral analysis, we excluded data where exposure-time loss was too significant to ensure reliable spectral statistics. The energy scale at 6 keV shows a slight negative shift in the high-count-rate regime. The energy resolution at 6 keV worsens as the count rate in electrically neighboring pixels increases, but can be restored by applying a nearest-neighbor coincidence cut (“cross-talk cut”). We examined how these effects influence the observation of bright point sources, using GX 13+1 as a test case, and identified an eV-scale energy offset at 6 keV between the inner (brighter) and outer (fainter) pixels. Users who seek to analyze velocity structures on the order of tens of km s–1 should account for such high-count-rate effects. These findings will aid in the interpretation of Resolve data from bright sources and provide valuable considerations for designing and planning for future microcalorimeter missions.

Publication Details

Journal
Maryland Shared Open Access Repository (USMAI Consortium)
Published
2025-07-11
DOI
https://doi.org/10.13016/m2w1bq-uvfg
Primary Topic
Food Security and Health in Diverse Populations
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

High-count-rate effects in event processing for the XRISM/Resolve X-ray microcalorimeter. II. Energy scale and resolution in orbit

Maryland Shared Open Access Repository (USMAI Consortium)
Food Security and Health in Diverse Populations
article

High-count-rate effects in event processing for the XRISM/Resolve X-ray microcalorimeter. II. Energy scale and resolution in orbit

article en

Abstract

The Resolve instrument on the X-ray Imaging and Spectroscopy Mission (XRISM) uses a 36 pixel microcalorimeter designed to deliver high-resolution, non-dispersive X-ray spectroscopy. Although it is optimized for extended sources with low count rates, Resolve observations of bright point sources are still able to provide unique insights into the physics of these objects, as long as high-count-rate effects are addressed in the analysis. These effects include the loss of exposure time for each pixel, changes in the energy scale, and changes in the energy resolution. To investigate these effects under realistic observational conditions, we observed the bright X-ray source, the Crab Nebula, with XRISM at several offset positions with respect to the Resolve field of view and with continuous illumination from 55Fe sources on the filter wheel. For the spectral analysis, we excluded data where exposure-time loss was too significant to ensure reliable spectral statistics. The energy scale at 6 keV shows a slight negative shift in the high-count-rate regime. The energy resolution at 6 keV worsens as the count rate in electrically neighboring pixels increases, but can be restored by applying a nearest-neighbor coincidence cut (“cross-talk cut”). We examined how these effects influence the observation of bright point sources, using GX 13+1 as a test case, and identified an eV-scale energy offset at 6 keV between the inner (brighter) and outer (fainter) pixels. Users who seek to analyze velocity structures on the order of tens of km s–1 should account for such high-count-rate effects. These findings will aid in the interpretation of Resolve data from bright sources and provide valuable considerations for designing and planning for future microcalorimeter missions.

Maryland Shared Open Access Repository (USMAI Consortium)
Affordable and clean energy
Openalex Percentile: Top 32%
Food Security and Health in Diverse Populations
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.