Hydrogen vs. Helium as carrier gases in GC-MS/MS for pesticide residue analysis: a comparative evaluation

• Hydrogen and helium compared as GC-MS/MS carrier gases for pesticide analysis • Equivalent method configurations used to ensure fair performance comparison • Helium showed better sensitivity at low levels, hydrogen enabled faster runs • Matrix effects were comparable across both gases with minor variations • Hydrogen proved viable for routine use, supporting greener lab practices This study presents a comprehensive comparison between helium and hydrogen as carrier gases in gas chromatography coupled with tandem mass spectrometry (GC-MS/MS) for multiresidue pesticide analysis. Using a shared instrumental platform, method parameters were carefully adapted for each carrier gas to maintain equivalent chromatographic efficiency, including adjustments to column dimensions, flow rates, and oven temperature programs. Method performance was evaluated in terms of chromatographic resolution, sensitivity, matrix effects, and regulatory reliability across tomato, pepper, and zucchini matrices. Hydrogen demonstrated faster analysis times and improved resolution in several cases, particularly in separating matrix interferences. However, helium consistently offered higher sensitivity, especially at lower concentrations. At 2 µg/kg, helium enabled the identification of over 90% of compounds in most matrices, while hydrogen remained below 55%. Matrix effect evaluations showed slightly better results with helium, though both gases provided acceptable performance, with most compounds falling within the negligible effect range. A proficiency test sample further validated the method, confirming that both gases produced results within acceptable z-score limits. Overall, hydrogen proved to be a viable and sustainable alternative for routine applications, though helium remains preferable when ultra-trace sensitivity or method robustness under stricter regulatory conditions is required.

Institutions

Publication Details

Journal
Green Analytical Chemistry
Published
2025-10-01
DOI
https://doi.org/10.1016/j.greeac.2025.100300
Primary Topic
Gut microbiota and health
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Hydrogen vs. Helium as carrier gases in GC-MS/MS for pesticide residue analysis: a comparative evaluation

Green Analytical Chemistry
Gut microbiota and health
article

Hydrogen vs. Helium as carrier gases in GC-MS/MS for pesticide residue analysis: a comparative evaluation

article en

Abstract

• Hydrogen and helium compared as GC-MS/MS carrier gases for pesticide analysis • Equivalent method configurations used to ensure fair performance comparison • Helium showed better sensitivity at low levels, hydrogen enabled faster runs • Matrix effects were comparable across both gases with minor variations • Hydrogen proved viable for routine use, supporting greener lab practices This study presents a comprehensive comparison between helium and hydrogen as carrier gases in gas chromatography coupled with tandem mass spectrometry (GC-MS/MS) for multiresidue pesticide analysis. Using a shared instrumental platform, method parameters were carefully adapted for each carrier gas to maintain equivalent chromatographic efficiency, including adjustments to column dimensions, flow rates, and oven temperature programs. Method performance was evaluated in terms of chromatographic resolution, sensitivity, matrix effects, and regulatory reliability across tomato, pepper, and zucchini matrices. Hydrogen demonstrated faster analysis times and improved resolution in several cases, particularly in separating matrix interferences. However, helium consistently offered higher sensitivity, especially at lower concentrations. At 2 µg/kg, helium enabled the identification of over 90% of compounds in most matrices, while hydrogen remained below 55%. Matrix effect evaluations showed slightly better results with helium, though both gases provided acceptable performance, with most compounds falling within the negligible effect range. A proficiency test sample further validated the method, confirming that both gases produced results within acceptable z-score limits. Overall, hydrogen proved to be a viable and sustainable alternative for routine applications, though helium remains preferable when ultra-trace sensitivity or method robustness under stricter regulatory conditions is required.

Green Analytical ChemistryVol. 15
European Union (BE)
Responsible consumption and production
Openalex Percentile: Top 48%
Gut microbiota and health
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.