Verification of analytical performance of cyclosporine assay on the Abbott Alinity ci® at the biochemistry laboratory of Mohammed VI university hospital Oujda

In medical laboratories, the verification of methods ensures precise measurements, complying with regulatory standards like the Moroccan Guide for Medical Laboratory Analysis (GBEA) and ISO 15189. After organ transplantation, blood levels of cyclosporine are often monitored closely, and its dosage adjusted accordingly to maintain therapeutic levels while minimizing the risk of toxicity. This study aims to verify the analytical performance of the cyclosporine assay on the Abbott Alinity ci® analyzer. Conducted at Mohammed VI university hospital’s biochemistry laboratory over 30 days, it involved two phases: evaluation of reproducibility and repeatability. Reproducibility was assessed by daily testing of the control samples at three concentration levels (low, medium, and high), and repeatability was measured by subjecting each serum sample to 30 individual assay runs. Cyclosporine determination used a dedicated reagent kit on the Abbott Alinity ci® analyzer’s immunology module, with data processing via the BYG middleware. The resulting coefficient of variation (CV) values were compared against standards set by the supplier. The investigation demonstrated a commendable repeatability across all concentration levels, with CV values of 3.59%, 3.23%, and 3.46% respectively. Intermediate fidelity examination yielded satisfactory results, with CV values of 3%, 4.83%, and 6.01% for low, medium, and high levels respectively. In conclusion, our study confirms the laboratory’s capability to provide accurate and precise cyclosporine assay results for clinical use, meeting the recommended criteria set by the suppliers.

Institutions

Publication Details

Journal
Zenodo (CERN European Organization for Nuclear Research)
Published
2025-03-30
DOI
https://doi.org/10.5281/zenodo.16930744
Primary Topic
Geochemistry and Geologic Mapping
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Verification of analytical performance of cyclosporine assay on the Abbott Alinity ci® at the biochemistry laboratory of Mohammed VI university hospital Oujda

Zenodo (CERN European Organization for Nuclear Research)
Geochemistry and Geologic Mapping
article

Verification of analytical performance of cyclosporine assay on the Abbott Alinity ci® at the biochemistry laboratory of Mohammed VI university hospital Oujda

article en

Abstract

In medical laboratories, the verification of methods ensures precise measurements, complying with regulatory standards like the Moroccan Guide for Medical Laboratory Analysis (GBEA) and ISO 15189. After organ transplantation, blood levels of cyclosporine are often monitored closely, and its dosage adjusted accordingly to maintain therapeutic levels while minimizing the risk of toxicity. This study aims to verify the analytical performance of the cyclosporine assay on the Abbott Alinity ci® analyzer. Conducted at Mohammed VI university hospital’s biochemistry laboratory over 30 days, it involved two phases: evaluation of reproducibility and repeatability. Reproducibility was assessed by daily testing of the control samples at three concentration levels (low, medium, and high), and repeatability was measured by subjecting each serum sample to 30 individual assay runs. Cyclosporine determination used a dedicated reagent kit on the Abbott Alinity ci® analyzer’s immunology module, with data processing via the BYG middleware. The resulting coefficient of variation (CV) values were compared against standards set by the supplier. The investigation demonstrated a commendable repeatability across all concentration levels, with CV values of 3.59%, 3.23%, and 3.46% respectively. Intermediate fidelity examination yielded satisfactory results, with CV values of 3%, 4.83%, and 6.01% for low, medium, and high levels respectively. In conclusion, our study confirms the laboratory’s capability to provide accurate and precise cyclosporine assay results for clinical use, meeting the recommended criteria set by the suppliers.

Zenodo (CERN European Organization for Nuclear Research)
Mohamed I University (MA)
Good health and well-being
Openalex Percentile: Top 68%
Geochemistry and Geologic Mapping
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.