Correlation between axial length measurements obtained from Aladdin optical biometer and axial length estimator

Background: Myopia is a significant public health concern associated with ocular pathologies like retinal detachment and glaucoma. Accurate measurement of axial length (AL) is crucial in myopia management. The Topcon Aladdin HW3.0 biometer is considered a gold standard, while the Axial Length Estimator (ALE) provides a cost-effective formula-based alternative. The ALE is a newer, more accessible tool for estimating axial length, using readily available clinical data like refractive error and corneal curvature. This study evaluates the correlation between the two methods and examines AL differences between genders. Methods: In this cross-sectional study, 99 participants underwent AL measurements using both the Topcon Aladdin HW3.0 and ALE formula. Statistical analysis included Pearson correlation, paired-sample t-tests, and independent-sample t-tests. Results: A strong positive correlation between the two methods was found (r = 0.853, p < 0.0005). However, a statistically significant difference was noted between the mean AL values (p = 0.032). Gender comparison yielded no significant difference in AL values using either method. Our findings suggest a strong correlation between the Topcon Aladdin HW3.0 biometer and the ALE. Despite this, the significant difference in mean AL values highlights potential limitations of the ALE, particularly in the precise measurement required for myopia management. The sample size may influence the lack of gender differences in AL. Conclusion: The ALE offers a promising alternative for AL measurement but is limited by significant differences from the biometer values, especially in clinical settings requiring precision. Further research is necessary to determine the ALE’s clinical applicability.

Publication Details

Journal
The International Islamic University Malaysia Repository (The International Islamic University Malaysia)
Published
2024-12-27
Primary Topic
Ophthalmology and Visual Impairment Studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Correlation between axial length measurements obtained from Aladdin optical biometer and axial length estimator

The International Islamic University Malaysia Repository (The International Islamic University Malaysia)
Ophthalmology and Visual Impairment Studies
article

Correlation between axial length measurements obtained from Aladdin optical biometer and axial length estimator

article en

Abstract

Background: Myopia is a significant public health concern associated with ocular pathologies like retinal detachment and glaucoma. Accurate measurement of axial length (AL) is crucial in myopia management. The Topcon Aladdin HW3.0 biometer is considered a gold standard, while the Axial Length Estimator (ALE) provides a cost-effective formula-based alternative. The ALE is a newer, more accessible tool for estimating axial length, using readily available clinical data like refractive error and corneal curvature. This study evaluates the correlation between the two methods and examines AL differences between genders. Methods: In this cross-sectional study, 99 participants underwent AL measurements using both the Topcon Aladdin HW3.0 and ALE formula. Statistical analysis included Pearson correlation, paired-sample t-tests, and independent-sample t-tests. Results: A strong positive correlation between the two methods was found (r = 0.853, p < 0.0005). However, a statistically significant difference was noted between the mean AL values (p = 0.032). Gender comparison yielded no significant difference in AL values using either method. Our findings suggest a strong correlation between the Topcon Aladdin HW3.0 biometer and the ALE. Despite this, the significant difference in mean AL values highlights potential limitations of the ALE, particularly in the precise measurement required for myopia management. The sample size may influence the lack of gender differences in AL. Conclusion: The ALE offers a promising alternative for AL measurement but is limited by significant differences from the biometer values, especially in clinical settings requiring precision. Further research is necessary to determine the ALE’s clinical applicability.

The International Islamic University Malaysia Repository (The International Islamic University Malaysia)
Gender equality
Openalex Percentile: Top 43%
Ophthalmology and Visual Impairment Studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Correlation between axial length measurements obtained from Aladdin optical biometer and axial length estimator · The International Islamic University Malaysia Repository (The International Islamic University Malaysia) (2024) | TGRS Research Map | TGRS