Reducing Measurement Error in Three-Point Bend Test Using Linear Gradient Correction Model

Endovascular devices like catheters feature non-uniform flexural rigidity (EI) profiles (flexi- ble tip, stiff shaft) crucial for navigation and pushability. Standard three-point bend analysis (assuming uniform stiffness) yields inaccurate apparent rigidity (B0) for these graded devices, especially in transition zones. This inaccuracy hinders device comparison, clinical selection, and predictive modeling. This thesis develops and validates a correction model for three-point bend tests aimed at improving local flexural rigidity accuracy for devices with varying stiffness, particularly those exhibiting transition zones similar to logistic profiles. The model assumes a linear gradient in rigidity (B(x) = Bm(1 + 2mx/L)) across the test span L. Solving the governing equation yields a correction factor W (m) relating true midpoint rigidity (Bm) to apparent rigidity via Bm = B0W (m). The gradient parameter m is estimated numerically via finite differences on sequential B0 measurements. The model’s performance was evaluated through numerical simulations using known true stiffness profiles. For ideal linear profiles, the correction accurately recovered the true stiffness (Bm ≈ Btrue), confirming the model’s theoretical validity. For non-linear logistic profiles, chosen as representative examples of catheter transitions, the correction significantly reduced viii percentage error compared to uncorrected B0, especially in high-gradient regions. Sensitivity analyses showed accuracy improves with smaller test spacing (better m estimation) but decreases for steeper profiles or profiles deviating significantly from the underlying linear assumption. Application to experimental catheter data demonstrated feasibility, yielding quantifiable corrections and a more representative local stiffness profile. In conclusion, the linear gradient model offers a computationally inexpensive method to potentially improve flexural rigidity characterization for devices whose stiffness profiles are reasonably approximated by the model’s assumptions (e.g., smooth, logistic-like transitions). While it can provide more reliable mechanical data under these conditions, aiding clinical decisions, device design, and modeling, its necessity and effectiveness may be limited for devices with more complex or abrupt stiffness changes.

Publication Details

Journal
Open Scholarship Institutional Repository (Washington University in St. Louis)
Published
2025-05-01
Primary Topic
Piezoelectric Actuators and Control
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Reducing Measurement Error in Three-Point Bend Test Using Linear Gradient Correction Model

Open Scholarship Institutional Repository (Washington University in St. Louis)
Piezoelectric Actuators and Control
article

Reducing Measurement Error in Three-Point Bend Test Using Linear Gradient Correction Model

article en

Abstract

Endovascular devices like catheters feature non-uniform flexural rigidity (EI) profiles (flexi- ble tip, stiff shaft) crucial for navigation and pushability. Standard three-point bend analysis (assuming uniform stiffness) yields inaccurate apparent rigidity (B0) for these graded devices, especially in transition zones. This inaccuracy hinders device comparison, clinical selection, and predictive modeling. This thesis develops and validates a correction model for three-point bend tests aimed at improving local flexural rigidity accuracy for devices with varying stiffness, particularly those exhibiting transition zones similar to logistic profiles. The model assumes a linear gradient in rigidity (B(x) = Bm(1 + 2mx/L)) across the test span L. Solving the governing equation yields a correction factor W (m) relating true midpoint rigidity (Bm) to apparent rigidity via Bm = B0W (m). The gradient parameter m is estimated numerically via finite differences on sequential B0 measurements. The model’s performance was evaluated through numerical simulations using known true stiffness profiles. For ideal linear profiles, the correction accurately recovered the true stiffness (Bm ≈ Btrue), confirming the model’s theoretical validity. For non-linear logistic profiles, chosen as representative examples of catheter transitions, the correction significantly reduced viii percentage error compared to uncorrected B0, especially in high-gradient regions. Sensitivity analyses showed accuracy improves with smaller test spacing (better m estimation) but decreases for steeper profiles or profiles deviating significantly from the underlying linear assumption. Application to experimental catheter data demonstrated feasibility, yielding quantifiable corrections and a more representative local stiffness profile. In conclusion, the linear gradient model offers a computationally inexpensive method to potentially improve flexural rigidity characterization for devices whose stiffness profiles are reasonably approximated by the model’s assumptions (e.g., smooth, logistic-like transitions). While it can provide more reliable mechanical data under these conditions, aiding clinical decisions, device design, and modeling, its necessity and effectiveness may be limited for devices with more complex or abrupt stiffness changes.

Open Scholarship Institutional Repository (Washington University in St. Louis)
Openalex Percentile: Top 45%
Piezoelectric Actuators and Control
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Reducing Measurement Error in Three-Point Bend Test Using Linear Gradient Correction Model · Open Scholarship Institutional Repository (Washington University in St. Louis) (2025) | TGRS Research Map | TGRS