The effectiveness of breast lead shielding on scattered radiation in lumbar spine radiography: a phantom study / Khairunnisa Adlina Hairul Anuar

This study evaluates the effectiveness of breast lead shielding in reducing scattered radiation in lumbar spine radiography. Using an anthropomorphic phantom model, an experimental study simulating anteroposterior (AP) and lateral lumbar spine radiography was performed with and without breast shielding. The scattered radiation doses were measured using Radcal ACCU-PRO dosimeter. Significant reductions in scattered radiation dose (42%- 49% reduction) were observed in both AP and lateral projections with the use of breast lead shielding as compared to without shielding (p < 0.001). These findings suggest that breast lead shielding effectively reduces radiation exposure in lumbar spine radiography. The study highlights the importance of using breast shielding to protect patients from scattered radiation. Further research should explore the long-term benefits of shielding, its application in other radiographic procedures and potential advancement in shielding materials and technique.

Publication Details

Journal
UiTM Institutional Repositories (Universiti Teknologi MARA)
Published
2024-10-01
Primary Topic
Study of Mite Species
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The effectiveness of breast lead shielding on scattered radiation in lumbar spine radiography: a phantom study / Khairunnisa Adlina Hairul Anuar

UiTM Institutional Repositories (Universiti Teknologi MARA)
Study of Mite Species
article

The effectiveness of breast lead shielding on scattered radiation in lumbar spine radiography: a phantom study / Khairunnisa Adlina Hairul Anuar

article en

Abstract

This study evaluates the effectiveness of breast lead shielding in reducing scattered radiation in lumbar spine radiography. Using an anthropomorphic phantom model, an experimental study simulating anteroposterior (AP) and lateral lumbar spine radiography was performed with and without breast shielding. The scattered radiation doses were measured using Radcal ACCU-PRO dosimeter. Significant reductions in scattered radiation dose (42%- 49% reduction) were observed in both AP and lateral projections with the use of breast lead shielding as compared to without shielding (p < 0.001). These findings suggest that breast lead shielding effectively reduces radiation exposure in lumbar spine radiography. The study highlights the importance of using breast shielding to protect patients from scattered radiation. Further research should explore the long-term benefits of shielding, its application in other radiographic procedures and potential advancement in shielding materials and technique.

UiTM Institutional Repositories (Universiti Teknologi MARA)
Openalex Percentile: Top 27%
Study of Mite Species
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.