Performance Evaluation of Suppression Structure for Superfluid Helium Film Flow at the Exit of Dilution Refrigerator Still Chamber
Superfluid helium (<sup>4</sup>He) exhibits special phenomena such as anomalous thermal conduction, fountain, and crawling film effects at sub-Kelvin temperatures owing to its quantum properties. Dilution refrigerators utilize <sup>3</sup>He as the working fluid for circulation. <sup>3</sup>He produces cooling effects as it passes through the phase interface in the mixing chamber. It is then pumped out by a room-temperature pump after evaporation in the still chamber. The superfluid helium inside the still chamber forms a liquid film on the wall and flows toward higher temperatures. This results in significant heat leakage owing to its anomalous thermal conductivity. The superfluid helium liquid film also affects the purity of <sup>3</sup>He during the circulation process, thereby influencing the cooling capacity of the dilution refrigerator. The “knife-edge” and “small orifice” structures are the primary means for suppressing superfluid helium films. This study analyzed the impact of the purity of the working fluid <sup>3</sup>He on the cooling capacity of the dilution refrigerator and the limitation placed on the system's <sup>3</sup>He circulation flow rate by the size of the “small orifice”. Experimental tests were conducted to measure the mass flow rate of the superfluid helium crawling film at 0.95 K with two types of suppression structures: “small orifice” and “knife-edge + small orifice”. The use of “knife-edge + small orifice” resulted in a 35.6% improvement in the suppression effect of the superfluid helium crawling film compared with the use of “small orifice” alone.
Publication Details
- Journal
- SHILAP Revista de lepidopterología
- Published
- 2025-01-01
- Primary Topic
- Quantum, superfluid, helium dynamics
- Type
- article
- Field-Weighted Citation Impact
- 0.00