Optimizing CT Radiation Exposure: A Comparative Study of CTDI and DLP in Head and Thorax Imaging

Computed Tomography (CT) is widely used in diagnostic medicine due to its speed and clarity, though it exposes patients to radiation. This study aimed to assess and compare the CTDI and DLP values in CT Head and CT Thorax scans using a Siemens 128-slice CT scanner at NIMS Hospital. A total of 200 adult patients were studied—100 for each scan type. The average CTDI and DLP for CT Head were found to be 56.93 mGy and 1113.84 mGy·cm respectively, while for CT Thorax, they were significantly lower at 8.44 mGy and 334.52 mGy·cm. The values were compared to national and international Diagnostic Reference Levels (DRLs), indicating that while CT Head values approached upper limits in some cases, they remained within global standards. CT Thorax values were consistently lower than both regional and global DRLs, reflecting effective dose optimization. These findings suggest that protocols for thoracic imaging at the center are well optimized, whereas CT Head protocols may benefit from further refinement. Overall, the study highlights the importance of continuous review of CT radiation exposure to ensure patient safety without compromising diagnostic quality.

Publication Details

Journal
Zenodo (CERN European Organization for Nuclear Research)
Published
2025-06-08
DOI
https://doi.org/10.5281/zenodo.15618256
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Optimizing CT Radiation Exposure: A Comparative Study of CTDI and DLP in Head and Thorax Imaging

Zenodo (CERN European Organization for Nuclear Research)
Radiation Dose and Imaging
article

Optimizing CT Radiation Exposure: A Comparative Study of CTDI and DLP in Head and Thorax Imaging

article en

Abstract

Computed Tomography (CT) is widely used in diagnostic medicine due to its speed and clarity, though it exposes patients to radiation. This study aimed to assess and compare the CTDI and DLP values in CT Head and CT Thorax scans using a Siemens 128-slice CT scanner at NIMS Hospital. A total of 200 adult patients were studied—100 for each scan type. The average CTDI and DLP for CT Head were found to be 56.93 mGy and 1113.84 mGy·cm respectively, while for CT Thorax, they were significantly lower at 8.44 mGy and 334.52 mGy·cm. The values were compared to national and international Diagnostic Reference Levels (DRLs), indicating that while CT Head values approached upper limits in some cases, they remained within global standards. CT Thorax values were consistently lower than both regional and global DRLs, reflecting effective dose optimization. These findings suggest that protocols for thoracic imaging at the center are well optimized, whereas CT Head protocols may benefit from further refinement. Overall, the study highlights the importance of continuous review of CT radiation exposure to ensure patient safety without compromising diagnostic quality.

Zenodo (CERN European Organization for Nuclear Research)
Openalex Percentile: Top 38%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.