Gray values and noise behavior of cone-beam computed tomography machines-an in vitro study.

Objectives To systematically evaluate the mean gray values (MGV) and noise provided by bone and soft tissue equivalent materials and air imaged with varied acquisition parameters in nine cone-beam computed tomography (CBCT) machines.Methods The DIN6868-161 phantom, composed of bone and soft tissue equivalent material and air gap, was scanned in nine CBCT machines. Tube current (mA) and tube voltage (kV), field of view (FOV) size, and rotation angle were varied over the possible range. The effect of the acquisition parameters on the MGV and contrast-to-noise indicator (CNI) was analyzed by Kruskal Wallis and Dunn-Bonferroni tests for each machine independently (α = 0.05).Results Tube current did not influence MGV in most machines. Viso G7 and Veraview X800 presented a decrease in the MGV for increasing kV. For ProMax 3D MAX and X1, the kV did not affect the MGV. For the majority of machines, MGV decreased with increasing FOV height. In general, the rotation angle did not affect the MGV. In addition, CNI was lower with lower radiation and large FOV and did not change from 80 kV in all machines.Conclusions The MGV and noise provided by the tested phantom vary largely among machines. The MGV is mainly influenced by the FOV size, especially for bone equivalent radiodensity. For most machines, when the acquisition parameters affect the MGV, the MGV decrease with the increase in the acquisition parameters.Advances In Knowledge Knowing the expected GV behavior in different exposure conditions hold potential for future calibration of MGV among CBCT machines.

Institutions

Publication Details

Journal
Open Access CRIS of the University of Bern
Published
2024-12-02
DOI
https://doi.org/10.48620/77086
Primary Topic
Dental Radiography and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Gray values and noise behavior of cone-beam computed tomography machines-an in vitro study.

Open Access CRIS of the University of Bern
Dental Radiography and Imaging
article

Gray values and noise behavior of cone-beam computed tomography machines-an in vitro study.

article en

Abstract

Objectives To systematically evaluate the mean gray values (MGV) and noise provided by bone and soft tissue equivalent materials and air imaged with varied acquisition parameters in nine cone-beam computed tomography (CBCT) machines.Methods The DIN6868-161 phantom, composed of bone and soft tissue equivalent material and air gap, was scanned in nine CBCT machines. Tube current (mA) and tube voltage (kV), field of view (FOV) size, and rotation angle were varied over the possible range. The effect of the acquisition parameters on the MGV and contrast-to-noise indicator (CNI) was analyzed by Kruskal Wallis and Dunn-Bonferroni tests for each machine independently (α = 0.05).Results Tube current did not influence MGV in most machines. Viso G7 and Veraview X800 presented a decrease in the MGV for increasing kV. For ProMax 3D MAX and X1, the kV did not affect the MGV. For the majority of machines, MGV decreased with increasing FOV height. In general, the rotation angle did not affect the MGV. In addition, CNI was lower with lower radiation and large FOV and did not change from 80 kV in all machines.Conclusions The MGV and noise provided by the tested phantom vary largely among machines. The MGV is mainly influenced by the FOV size, especially for bone equivalent radiodensity. For most machines, when the acquisition parameters affect the MGV, the MGV decrease with the increase in the acquisition parameters.Advances In Knowledge Knowing the expected GV behavior in different exposure conditions hold potential for future calibration of MGV among CBCT machines.

Open Access CRIS of the University of Bern
Stomatology Hospital (CN)
Openalex Percentile: Top 49%
Dental Radiography and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.